Recall Z-0708-2021— ICU Medical, Inc.
Class II · Ongoing
Class II FDA medical device recall by ICU Medical, Inc., initiated 2020-08-31. FDA's recall record names no specific 510(k) clearance for this event. Product: 76 cm (30") Appx 3.3 ml, Admin Set, 2 Spiros w/Red Cap, 20 Drop In-Line Drip Chamber w/15 Micron Filter, Bag Hanger, REF: 011-CH3261; 44 cm(17 IN) Appx 2.9 ml,PUR YELLOW TRANS SET,ChemoClave,CK VLV,Spiros,PURPLE CAP, REF: 011-CH3568; Oncology Kit, 5 Gang Stopcock w/5 ChemoClave, Spiros w/Red Cap; 10cm Smallbore Bifuse Ext w/2 ChemoClave, Red Ring, BCV; 297cm Ext w/ChemoClave, REF: 011-CH3600; Oncology Kit, 5 Gang Stopcock w/5 ChemoClave, Spiros w/Red Cap; 10cm Smallbore Bifuse Ext, w/2 ChemoClave, Red Ring, BCV; 597 cm Ext w/ChemoClave, REF: 011-CH3601; Kit, 2 5 Gang Stopcocks w/5 MicroClave Clear, Spiros; 2 Smallbore Bifuse Ext Sets, 2 MicroClave Clear, Red Ring, Check Valve, RMLL; 2 Ext Sets, MicroClave Clear Clipped Together, RMLL, REF: 011-CH3610; Admin Set w/Clave, Spiros, REF: 011-CH3772; Admin Set w/Clave, Spiros, REF: 011-CH3773; Admin Set w/Clave, Spiros, REF: 011-CH3967; 76" (193 cm) 50 ml, Diluent Set for Channel 2 w/20 Drop Drip Chamber, Spiros, Drop-In Red Cap, REF: 011-CH4011. Reason: Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line..
- Recalling firm
- ICU Medical, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2020-08-31
- Firm location
- San Clemente, CA, United States
Product description
76 cm (30") Appx 3.3 ml, Admin Set, 2 Spiros w/Red Cap, 20 Drop In-Line Drip Chamber w/15 Micron Filter, Bag Hanger, REF: 011-CH3261; 44 cm(17 IN) Appx 2.9 ml,PUR YELLOW TRANS SET,ChemoClave,CK VLV,Spiros,PURPLE CAP, REF: 011-CH3568; Oncology Kit, 5 Gang Stopcock w/5 ChemoClave, Spiros w/Red Cap; 10cm Smallbore Bifuse Ext w/2 ChemoClave, Red Ring, BCV; 297cm Ext w/ChemoClave, REF: 011-CH3600; Oncology Kit, 5 Gang Stopcock w/5 ChemoClave, Spiros w/Red Cap; 10cm Smallbore Bifuse Ext, w/2 ChemoClave, Red Ring, BCV; 597 cm Ext w/ChemoClave, REF: 011-CH3601; Kit, 2 5 Gang Stopcocks w/5 MicroClave Clear, Spiros; 2 Smallbore Bifuse Ext Sets, 2 MicroClave Clear, Red Ring, Check Valve, RMLL; 2 Ext Sets, MicroClave Clear Clipped Together, RMLL, REF: 011-CH3610; Admin Set w/Clave, Spiros, REF: 011-CH3772; Admin Set w/Clave, Spiros, REF: 011-CH3773; Admin Set w/Clave, Spiros, REF: 011-CH3967; 76" (193 cm) 50 ml, Diluent Set for Channel 2 w/20 Drop Drip Chamber, Spiros, Drop-In Red Cap, REF: 011-CH4011
Reason for recall
Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
Data sourced from openFDA. This site is unofficial and independent of the FDA.