Recall Z-0707-2017— Merge Healthcare, Inc.
Class II · Terminated
Class II FDA medical device recall by Merge Healthcare, Inc., initiated 2016-10-20, terminated 2019-05-30. It names one FDA 510(k) clearance: K020995. Product: Merge, eFilm Workstation and eFilm Lite eFilm Workstation with Modules is a software application that is used for viewing medical images. eFilm Workstation with Modules receives digital images and data from various sources (including but not limited to CT, MR, US, RF units, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways or imaging sources). eFilm Workstation with Modules can be used to communicate, process and display medical images. Users have access to various image processing and measurement tools to assist them in viewing images. In addition, users can overlay templates on medical images to aid in preoperative planning. eFilm Workstation with Modules can be integrated with an institution s existing HIS or RIS for a fully integrated electronic patient record. Typical users of eFilm Workstation with Modules are trained medical professionals, including but not limited to radiologists, technologists and clinicians. Reason: A product issues happens for RF projection images. If user measures on RF image, .cal (Calibration) is not displayed and measurement is not correct. The .cal label is not displayed on the projection images after calibration is performed which should be present if the measurement is presented based on Pixel Spacing..
- Recalling firm
- Merge Healthcare, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-10-20
- Terminated
- 2019-05-30
- Firm location
- Hartland, WI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Merge, eFilm Workstation and eFilm Lite eFilm Workstation with Modules is a software application that is used for viewing medical images. eFilm Workstation with Modules receives digital images and data from various sources (including but not limited to CT, MR, US, RF units, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways or imaging sources). eFilm Workstation with Modules can be used to communicate, process and display medical images. Users have access to various image processing and measurement tools to assist them in viewing images. In addition, users can overlay templates on medical images to aid in preoperative planning. eFilm Workstation with Modules can be integrated with an institution s existing HIS or RIS for a fully integrated electronic patient record. Typical users of eFilm Workstation with Modules are trained medical professionals, including but not limited to radiologists, technologists and clinicians
Reason for recall
A product issues happens for RF projection images. If user measures on RF image, .cal (Calibration) is not displayed and measurement is not correct. The .cal label is not displayed on the projection images after calibration is performed which should be present if the measurement is presented based on Pixel Spacing.