Recall Z-0706-2021— ICU Medical, Inc.

Class II · Ongoing

Class II FDA medical device recall by ICU Medical, Inc., initiated 2020-08-31. FDA's recall record names no specific 510(k) clearance for this event. Product: Spiros CLOSED MALE LUER w/RED CAP, 25 UNITS, REF: 011-CH2000SC-25; 76 cm (30") Appx 3.2 ml, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, Spiros w/Purple Cap, Bag Hanger, REF: 011-CH3185; 46 cm (18") Appx 7.4 ml, Trifuse Add-On Set w/2 Spiros w/Purple Caps, Bag Spike, 3 Clamps (Blue, 2 Red), Dry Spike Adapter, REF: 011-CH3266; 76 cm (30") Appx 3.3 ml, Admin Set, 2 Spiros w/Purple Cap, 20 Drop In-Line Drip Chamber, Bag Hanger, REF: 011-CH3446; 61 cm (24") Appx 8.2 ml, Bifuse Add-On Set w/Bag Spike, Spiros w/Purple Cap, 3 Clamps (1 Blue, 2 White), Check Valve, Dry Spike Adapter, REF: 011-CH3729; 25 UNITS of Spiros CLOSED MALE LUER, REF: 011-H2457; Appx 1.0 ml, Adattatore per Pompa con Spiros, REF: 011-H2629; 179 cm (70") Appx 11.5 ml, PUR Ambr. Deflussore con Spiros, Camera, BCV, Y-Clave e LL Gir., REF: 011-H2776; 180 cm (71IN)APPX 11.5ml,PUR AMBR,Spiros,CAMERA,REG.FLUSSO,BCV,Y-Clave, REF: 011-H2864-HSR; 181cm(71 IN)APPX11.6ml,PUR DEFLUSSORE,Spiros,20 DROP IN LINE,BCV,Clave, REF: 011-H2904; 38 cm (15 IN) APPX 2.9 ml, TRIFUSE EXT SET, 3 MicroClave, Spiros, PURPLE CAP, REF: 011-H3449; 42cm(16.5 IN)APPX 2.6ml,PUR TRANSFER SET,Clave ADDITIVE PORT,BCV,Spiros, REF: 011-H3597; 41cm(16 IN)APPX 2.9ml,PUR SET PERFORATORE,Clave,VALVOLA,Spiros, REF: 011-H3628; 41cm(16 IN)APPX 3.1ml,SET AMBRATO,PERFORATORE,Clave,VALVOLA,Spiros, REF: 011-H3629; 42cm(16.5 IN)APPX 5.6ml,SET,PERFORATORE,Clave,FILTRO,VALVOLA,Spiros, REF: 011-H3630; 109 cm(43 IN) APPX 7.2 ml,KIT,EXT SET W/Spiros,STOPCOCK,5 Clave,4-GANG STOPCOCK, REF: 011-MC330632; 2 Ext Sets w/5-Gang Stopcock, 9 Clave, Spiros, Stopcock, REF: 011-MC33242; Kit for multiple transfers, REF: 011-N1000; Kit per il riempimento pompe elastomeriche con SPIROS / Kit for elastomeric pump priming with SPIROS, REF: 011-N1000-P2; APPX 2.3ml, ZYTO-VERBINDUNGSLEITUNG (ZYTO-CONNECTING LINE), REF: 011-N1664SP; Appx 1.0 ml, Adaptador universal para sistemas IV con Spiros, REF: 034-H2629; Spinning Spiros Closed Male Luer, Red Cap (No Flange), REF: CH2000S-NR; 15" (38 cm) Appx 5.2 ml, Bifuse Add-On Set w/Bag Spike, Spiros, 3 Clamps (Blue, 2 White), Dry Spike Adapter, REF: CH3187; 14" (36 cm) Appx 8.0 ml, Quinfuse Add-On Set w/4 Spiros, Bag Spike, 5 Clamps (Blue, 4 Red), Dry Spike Adapter, REF: CH3225; 20" (51 cm) Appx 9.5 ml, Quadfuse Add-On Set w/3 Spiros, Bag Spike, 4 Clamps (Blue, 3 Red), Dry Spike Adapter, REF: CH3226; 16" (41 cm) Appx 6.3 ml, Trifuse Add-On Set w/2 Spiros, Bag Spike, 3 Clamps (Blue, 2 Red), Dry Spike Adapter, REF: CH3227; 30" (76 cm) Appx 4.1 ml, Yellow 20 Drop Admin Set w/15 Micron Filter, Spiros, REF: CH3235; 28" (71 cm) Appx 5.1 ml, Add On Set w/Clave Spike, Spiros, Clamp, REF: CH3354; 16cm(6 IN)APPX 2.4ml,SET,MicroClave Clear,STOPCOCK,DROP-IN ADD-ON,Spiros, REF CH3411: APPX 1.9 ml,13 cm (5 IN),BAG SPIKE,Spiros,ACCESS DEVICE w/Clave, REF: CH3502; 60 (152 cm) Appx 1.9 ml, Smallbore Ext Set w/MicroClave Clear, 0.2 Micron Low Protein Binding Filter, Spiros w/Red Cap, Clamp, REF: CH3603; Spinning Spiros Closed Male Luer, REF: K7040-001; APPX 0.76ml, 4-WAY STOPCOCK w/2 Clave, VIAL ADAPTER, 3 Spiros, REF: K7052-001; ZYTOmed Verab.-Set Onko II lichtschutz, REF: SN1205AG; ZYTOmed - Verabreichungs-Set Onko III lichtgesch¿tzt, REF: SN1206A; Conector macho cerrado Spinning Spiros, tapon purpura, REF: 034-CH2000S-PC. Reason: Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line..

Recalling firm
ICU Medical, Inc.
Classification
Class II
Status
Ongoing
Initiated
2020-08-31
Firm location
San Clemente, CA, United States

Product description

Spiros CLOSED MALE LUER w/RED CAP, 25 UNITS, REF: 011-CH2000SC-25; 76 cm (30") Appx 3.2 ml, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, Spiros w/Purple Cap, Bag Hanger, REF: 011-CH3185; 46 cm (18") Appx 7.4 ml, Trifuse Add-On Set w/2 Spiros w/Purple Caps, Bag Spike, 3 Clamps (Blue, 2 Red), Dry Spike Adapter, REF: 011-CH3266; 76 cm (30") Appx 3.3 ml, Admin Set, 2 Spiros w/Purple Cap, 20 Drop In-Line Drip Chamber, Bag Hanger, REF: 011-CH3446; 61 cm (24") Appx 8.2 ml, Bifuse Add-On Set w/Bag Spike, Spiros w/Purple Cap, 3 Clamps (1 Blue, 2 White), Check Valve, Dry Spike Adapter, REF: 011-CH3729; 25 UNITS of Spiros CLOSED MALE LUER, REF: 011-H2457; Appx 1.0 ml, Adattatore per Pompa con Spiros, REF: 011-H2629; 179 cm (70") Appx 11.5 ml, PUR Ambr. Deflussore con Spiros, Camera, BCV, Y-Clave e LL Gir., REF: 011-H2776; 180 cm (71IN)APPX 11.5ml,PUR AMBR,Spiros,CAMERA,REG.FLUSSO,BCV,Y-Clave, REF: 011-H2864-HSR; 181cm(71 IN)APPX11.6ml,PUR DEFLUSSORE,Spiros,20 DROP IN LINE,BCV,Clave, REF: 011-H2904; 38 cm (15 IN) APPX 2.9 ml, TRIFUSE EXT SET, 3 MicroClave, Spiros, PURPLE CAP, REF: 011-H3449; 42cm(16.5 IN)APPX 2.6ml,PUR TRANSFER SET,Clave ADDITIVE PORT,BCV,Spiros, REF: 011-H3597; 41cm(16 IN)APPX 2.9ml,PUR SET PERFORATORE,Clave,VALVOLA,Spiros, REF: 011-H3628; 41cm(16 IN)APPX 3.1ml,SET AMBRATO,PERFORATORE,Clave,VALVOLA,Spiros, REF: 011-H3629; 42cm(16.5 IN)APPX 5.6ml,SET,PERFORATORE,Clave,FILTRO,VALVOLA,Spiros, REF: 011-H3630; 109 cm(43 IN) APPX 7.2 ml,KIT,EXT SET W/Spiros,STOPCOCK,5 Clave,4-GANG STOPCOCK, REF: 011-MC330632; 2 Ext Sets w/5-Gang Stopcock, 9 Clave, Spiros, Stopcock, REF: 011-MC33242; Kit for multiple transfers, REF: 011-N1000; Kit per il riempimento pompe elastomeriche con SPIROS / Kit for elastomeric pump priming with SPIROS, REF: 011-N1000-P2; APPX 2.3ml, ZYTO-VERBINDUNGSLEITUNG (ZYTO-CONNECTING LINE), REF: 011-N1664SP; Appx 1.0 ml, Adaptador universal para sistemas IV con Spiros, REF: 034-H2629; Spinning Spiros Closed Male Luer, Red Cap (No Flange), REF: CH2000S-NR; 15" (38 cm) Appx 5.2 ml, Bifuse Add-On Set w/Bag Spike, Spiros, 3 Clamps (Blue, 2 White), Dry Spike Adapter, REF: CH3187; 14" (36 cm) Appx 8.0 ml, Quinfuse Add-On Set w/4 Spiros, Bag Spike, 5 Clamps (Blue, 4 Red), Dry Spike Adapter, REF: CH3225; 20" (51 cm) Appx 9.5 ml, Quadfuse Add-On Set w/3 Spiros, Bag Spike, 4 Clamps (Blue, 3 Red), Dry Spike Adapter, REF: CH3226; 16" (41 cm) Appx 6.3 ml, Trifuse Add-On Set w/2 Spiros, Bag Spike, 3 Clamps (Blue, 2 Red), Dry Spike Adapter, REF: CH3227; 30" (76 cm) Appx 4.1 ml, Yellow 20 Drop Admin Set w/15 Micron Filter, Spiros, REF: CH3235; 28" (71 cm) Appx 5.1 ml, Add On Set w/Clave Spike, Spiros, Clamp, REF: CH3354; 16cm(6 IN)APPX 2.4ml,SET,MicroClave Clear,STOPCOCK,DROP-IN ADD-ON,Spiros, REF CH3411: APPX 1.9 ml,13 cm (5 IN),BAG SPIKE,Spiros,ACCESS DEVICE w/Clave, REF: CH3502; 60 (152 cm) Appx 1.9 ml, Smallbore Ext Set w/MicroClave Clear, 0.2 Micron Low Protein Binding Filter, Spiros w/Red Cap, Clamp, REF: CH3603; Spinning Spiros Closed Male Luer, REF: K7040-001; APPX 0.76ml, 4-WAY STOPCOCK w/2 Clave, VIAL ADAPTER, 3 Spiros, REF: K7052-001; ZYTOmed Verab.-Set Onko II lichtschutz, REF: SN1205AG; ZYTOmed - Verabreichungs-Set Onko III lichtgesch¿tzt, REF: SN1206A; Conector macho cerrado Spinning Spiros, tapon purpura, REF: 034-CH2000S-PC

Reason for recall

Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.

Data sourced from openFDA. This site is unofficial and independent of the FDA.