Recall Z-0705-2025— GE OEC Medical Systems, Inc

Class II · Ongoing

Class II FDA medical device recall by GE OEC Medical Systems, Inc, initiated 2024-11-21. It names 4 FDA 510(k) clearances: K122234, K120613, K082781 and 1 more. Product: OEC 9900 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 9900 Elite mobile fluoroscopy system is designed to provide fluoroscopic and spot-film images of the patient during diagnostic surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures.. Reason: Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability..

Recalling firm
GE OEC Medical Systems, Inc
Classification
Class II
Status
Ongoing
Initiated
2024-11-21
Firm location
Salt Lake City, UT, United States

510(k) clearances named in this recall

4 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

OEC 9900 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 9900 Elite mobile fluoroscopy system is designed to provide fluoroscopic and spot-film images of the patient during diagnostic surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures.

Reason for recall

Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.

Data sourced from openFDA. This site is unofficial and independent of the FDA.