Recall Z-0703-2013— GE Healthcare, LLC

Class II · Terminated

Class II FDA medical device recall by GE Healthcare, LLC, initiated 2012-09-26, terminated 2013-12-05. FDA's recall record names no specific 510(k) clearance for this event. Product: GE Healthcare Definium 5000 / 5220493 X-Ray System. Reason: When quickly switching the protocols from AEC to Fixed under the same view, if the user selection process time is less than 0.5 seconds, the User Interface (UI) may display different exposure techniques than selected, and may not display AEC setting. The exposure can still be made as commanded, however it may not be properly indicated..

Recalling firm
GE Healthcare, LLC
Classification
Class II
Status
Terminated
Initiated
2012-09-26
Terminated
2013-12-05
Firm location
Waukesha, WI, United States

Product description

GE Healthcare Definium 5000 / 5220493 X-Ray System

Reason for recall

When quickly switching the protocols from AEC to Fixed under the same view, if the user selection process time is less than 0.5 seconds, the User Interface (UI) may display different exposure techniques than selected, and may not display AEC setting. The exposure can still be made as commanded, however it may not be properly indicated.