Recall Z-0702-2013— Philips Healthcare Inc.
Class II · Terminated
Class II FDA medical device recall by Philips Healthcare Inc., initiated 2012-12-13, terminated 2017-01-24. It names one FDA 510(k) clearance: K050190. Product: Juno DRF , X-ray system, diagnostic, Fluoroscopic, general-purpose Model: 70920 Product Usage: Universal R/F, general use. Reason: Values of fluoroscopy time, Air Kerma Rate and Cumulated Air Kerma may not be displayed "at the fluoroscopist's working position" as per 21 CFR 1003.21.
- Recalling firm
- Philips Healthcare Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2012-12-13
- Terminated
- 2017-01-24
- Firm location
- Andover, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Juno DRF , X-ray system, diagnostic, Fluoroscopic, general-purpose Model: 70920 Product Usage: Universal R/F, general use
Reason for recall
Values of fluoroscopy time, Air Kerma Rate and Cumulated Air Kerma may not be displayed "at the fluoroscopist's working position" as per 21 CFR 1003.21
Data sourced from openFDA. This site is unofficial and independent of the FDA.