Recall Z-0701-2021— ICU Medical, Inc.

Class II · Ongoing

Class II FDA medical device recall by ICU Medical, Inc., initiated 2020-08-31. It names one FDA 510(k) clearance: K081361. Product: 5 IN (13 cm) Appx 1.5 ml, Bag Spike Adapter, Spiros, REF: 20123-01; 75" (191 cm) Appx 9.1 ml, 20 Drop Admin Set w/ Integrated Clave Drip Chamber, Clave, Spiros, REF: CH3035; 40" (102 cm) Appx 4.5 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, Spiros, Hanger, REF: CH3135; 14" (36 cm) Appx 4.8 ml, Bifuse Add-On Set w/Spiros w/Red Cap, Bag Spike, 3 Clamps (Blue, 2 White), Vented Cap, Drop-In Spiros w/Red Cap, REF: CH3156; 114" Bifuse Primary Set w/2 Spiros, Bag Spike, 2 20 Drop In-line Drip Chamber, Clave, Bag Hanger, REF: CH3240; 18 IN (46 cm) Appx 2.9 ml, Add-On Set w/Spiros, Clave Spike, Clamp, Cap, REF: CH3290; 27" (69 cm) Appx 5.2 ml, Bifuse Add-On Set w/15 Drop Drip Chamber, Spiros w/Red Cap, Check Valve, Vented Cap, Bag Hanger, REF: CH3294; Secondary IV Oncology Kit, REF: CH3322; 30 IN (76 cm) Appx 1.8 ml) Ext Set w/2 Spiros, 2 Red Caps, Clamp, REF: CH3327; 40" (102 cm) Appx 4.6 ml, Admin Set, 2 Spiros w/2 Red Caps, 20 Drop In-Line Drip Chamber, 2 Bag Hangers, REF: CH3373; Oncology Kit w/Spiros, Red Cap, Bag Spike w/Clave Additive Port, REF: CH3397; 45" (114 cm) Appx 5.8 ml, Ext Set w/MicroClave, 2 Clave, Spiros w/Red Cap, 2 Clamps, REF: CH3403; Oncology Kit w/20" (51 cm) Ext Set w/MicroClave, Graduated Connector, Transfer Set w/Clave Additive Port, Spiros, Red Cap, 13mm Vial Spike w/Clave, REF: CH3507; Oncology Kit w/Clave Universal Vented Vial Spike, Spinning Spiros w/Red Cap, REF: CH3546; 22" (56 cm) Add-On 150 ml Burette Set (MicroClave, No Shut-Off), w/Spiros, 2 Clamps, Vented Cap, REF: CH3554; 4.5 IN (11 cm) Appx 0.43 ml, Ext Set, Spiros w/Red Cap, REF: CH3559; Oncology Kit w/Spinning Spiros, C-Clip, Red Cap, REF: CH3581; 5 IN (13 cm) Appx 0.57 ml, Ext Set, Anti-Siphon Valve, Spiros, REF: CH3582; Oncology Kit w/Spinning Spiros w/Red Cap, C-Clip; Priming Cap, REF: CH3589; 77" (196 cm) Appx. 9.4 ml, Admin Set w/2 Spiros w/Red Cap, 20 Drop In-Line Drip Chamber, Bag Hanger, REF: CH3639; 22" (56 cm) Appx 5.6ml, Ext Set, Trifuse w/4 Clave Clear, Spiros w/Red Cap, 1.2 Micron Filter, 4 Clamps, REF: CH3770; 6.5 IN (17 cm) Appx 0.78 ml, Ext Set w/Clave Clear, Clave 4-Way Stopcock, Spiros, Red Cap, REF: CH3771; 47" (119 cm) 50 ml Diluent Set for Channel 2 w/20 Drop In-Line Drip Chamber, 2 Spiros, REF: CH4001; 47" (119 cm) 50 ml Diluent Set For Channel 2 w/Bag Spike w/MicroClave Additive Port, Spiros, REF: CH4009; 30" (76 cm) Appx 5.9 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, 0.2 Micron Filter, Spiros, Hanger, REF: Z7100; 30" (76 cm) 50 ml Diluent Set for Channel 2 w/Spiros, REF: CH4006; 60 IN(152cm)APPX 0.45ml,SMALLBORE EXT,MicroCLAVE,Spiros, REF: Z7040; 6" (15 cm) Appx 0.55 ml, Smallbore Trifuse Ext Set w/3 MicroClave, 3 Clamps, Spiros, REF: Z7151. Reason: Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line..

Recalling firm
ICU Medical, Inc.
Classification
Class II
Status
Ongoing
Initiated
2020-08-31
Firm location
San Clemente, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

5 IN (13 cm) Appx 1.5 ml, Bag Spike Adapter, Spiros, REF: 20123-01; 75" (191 cm) Appx 9.1 ml, 20 Drop Admin Set w/ Integrated Clave Drip Chamber, Clave, Spiros, REF: CH3035; 40" (102 cm) Appx 4.5 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, Spiros, Hanger, REF: CH3135; 14" (36 cm) Appx 4.8 ml, Bifuse Add-On Set w/Spiros w/Red Cap, Bag Spike, 3 Clamps (Blue, 2 White), Vented Cap, Drop-In Spiros w/Red Cap, REF: CH3156; 114" Bifuse Primary Set w/2 Spiros, Bag Spike, 2 20 Drop In-line Drip Chamber, Clave, Bag Hanger, REF: CH3240; 18 IN (46 cm) Appx 2.9 ml, Add-On Set w/Spiros, Clave Spike, Clamp, Cap, REF: CH3290; 27" (69 cm) Appx 5.2 ml, Bifuse Add-On Set w/15 Drop Drip Chamber, Spiros w/Red Cap, Check Valve, Vented Cap, Bag Hanger, REF: CH3294; Secondary IV Oncology Kit, REF: CH3322; 30 IN (76 cm) Appx 1.8 ml) Ext Set w/2 Spiros, 2 Red Caps, Clamp, REF: CH3327; 40" (102 cm) Appx 4.6 ml, Admin Set, 2 Spiros w/2 Red Caps, 20 Drop In-Line Drip Chamber, 2 Bag Hangers, REF: CH3373; Oncology Kit w/Spiros, Red Cap, Bag Spike w/Clave Additive Port, REF: CH3397; 45" (114 cm) Appx 5.8 ml, Ext Set w/MicroClave, 2 Clave, Spiros w/Red Cap, 2 Clamps, REF: CH3403; Oncology Kit w/20" (51 cm) Ext Set w/MicroClave, Graduated Connector, Transfer Set w/Clave Additive Port, Spiros, Red Cap, 13mm Vial Spike w/Clave, REF: CH3507; Oncology Kit w/Clave Universal Vented Vial Spike, Spinning Spiros w/Red Cap, REF: CH3546; 22" (56 cm) Add-On 150 ml Burette Set (MicroClave, No Shut-Off), w/Spiros, 2 Clamps, Vented Cap, REF: CH3554; 4.5 IN (11 cm) Appx 0.43 ml, Ext Set, Spiros w/Red Cap, REF: CH3559; Oncology Kit w/Spinning Spiros, C-Clip, Red Cap, REF: CH3581; 5 IN (13 cm) Appx 0.57 ml, Ext Set, Anti-Siphon Valve, Spiros, REF: CH3582; Oncology Kit w/Spinning Spiros w/Red Cap, C-Clip; Priming Cap, REF: CH3589; 77" (196 cm) Appx. 9.4 ml, Admin Set w/2 Spiros w/Red Cap, 20 Drop In-Line Drip Chamber, Bag Hanger, REF: CH3639; 22" (56 cm) Appx 5.6ml, Ext Set, Trifuse w/4 Clave Clear, Spiros w/Red Cap, 1.2 Micron Filter, 4 Clamps, REF: CH3770; 6.5 IN (17 cm) Appx 0.78 ml, Ext Set w/Clave Clear, Clave 4-Way Stopcock, Spiros, Red Cap, REF: CH3771; 47" (119 cm) 50 ml Diluent Set for Channel 2 w/20 Drop In-Line Drip Chamber, 2 Spiros, REF: CH4001; 47" (119 cm) 50 ml Diluent Set For Channel 2 w/Bag Spike w/MicroClave Additive Port, Spiros, REF: CH4009; 30" (76 cm) Appx 5.9 ml, 20 Drop Admin Set w/Integrated Clave Drip Chamber, 0.2 Micron Filter, Spiros, Hanger, REF: Z7100; 30" (76 cm) 50 ml Diluent Set for Channel 2 w/Spiros, REF: CH4006; 60 IN(152cm)APPX 0.45ml,SMALLBORE EXT,MicroCLAVE,Spiros, REF: Z7040; 6" (15 cm) Appx 0.55 ml, Smallbore Trifuse Ext Set w/3 MicroClave, 3 Clamps, Spiros, REF: Z7151

Reason for recall

Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.