Class II FDA medical device recall by Medtronic Neuromodulation, initiated 2016-10-03, terminated 2022-05-04. FDA's recall record names no specific 510(k) clearance for this event. Product: Model 37751 Recharger Product Usage: The Medtronic 37751 Recharger is a hand-held device used to charge the battery in a patient s neurostimulator. It includes a display to provide information on the charging system. The charging system consists of the Model 37751 Recharger, an AC power supply and power cord (Model 37761) and an antenna.. Reason: Medtronic has identified an increased number of complaints from customers involving reports of Rechargers (Medtronic Model 37751 Recharger, which is included in Models 37754 and 97754 Charging Systems,Spinal Cord Stimulation)that are in an unresponsive error state, where the Recharger is non-functional with a blank display screen and is beeping every 5 seconds. Medtronic has determined all Rechargers manufactured starting in November 2014 (indicated by serial numbers beginning with NKA4 or NKU4) are more susceptible to this error state..

Recalling firm
Medtronic Neuromodulation
Classification
Class II
Status
Terminated
Initiated
2016-10-03
Terminated
2022-05-04
Firm location
Minneapolis, MN, United States

Product description

Model 37751 Recharger Product Usage: The Medtronic 37751 Recharger is a hand-held device used to charge the battery in a patient s neurostimulator. It includes a display to provide information on the charging system. The charging system consists of the Model 37751 Recharger, an AC power supply and power cord (Model 37761) and an antenna.

Reason for recall

Medtronic has identified an increased number of complaints from customers involving reports of Rechargers (Medtronic Model 37751 Recharger, which is included in Models 37754 and 97754 Charging Systems,Spinal Cord Stimulation)that are in an unresponsive error state, where the Recharger is non-functional with a blank display screen and is beeping every 5 seconds. Medtronic has determined all Rechargers manufactured starting in November 2014 (indicated by serial numbers beginning with NKA4 or NKU4) are more susceptible to this error state.

Data sourced from openFDA. This site is unofficial and independent of the FDA.