Recall Z-0699-2013— Becton Dickinson & Co.
Class II · Terminated
Class II FDA medical device recall by Becton Dickinson & Co., initiated 2012-12-13, terminated 2013-06-03. It names one FDA 510(k) clearance: K926018. Product: Becton, Dickinson and Company, BBL Crystal Enteric/Nonfermenter ID Kit, Catalog # 24500, cartons of 20 kits, In-vitro diagnostic reagents for microbiological testing. The BBL Crystal Enteric/Nonfermentor (E/NF) identification (ID) System is for the identification of aerobic gram-negative bacteria that belong to the family Enterobacteriaceae as well as some of the more frequently isolated glucose fermenting and nonfermenting gram-negative bacilli.. Reason: In-vitro diagnostic reagents for microbiological testing may be reversed in well positions in the test kit and when utilized may lead to bacteria misidentification..
- Recalling firm
- Becton Dickinson & Co.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2012-12-13
- Terminated
- 2013-06-03
- Firm location
- Sparks, MD, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Becton, Dickinson and Company, BBL Crystal Enteric/Nonfermenter ID Kit, Catalog # 24500, cartons of 20 kits, In-vitro diagnostic reagents for microbiological testing. The BBL Crystal Enteric/Nonfermentor (E/NF) identification (ID) System is for the identification of aerobic gram-negative bacteria that belong to the family Enterobacteriaceae as well as some of the more frequently isolated glucose fermenting and nonfermenting gram-negative bacilli.
Reason for recall
In-vitro diagnostic reagents for microbiological testing may be reversed in well positions in the test kit and when utilized may lead to bacteria misidentification.