Recall Z-0695-2025— MEDLINE INDUSTRIES, LP - Northfield

Class II · Ongoing

Class II FDA medical device recall by MEDLINE INDUSTRIES, LP - Northfield, initiated 2024-11-18. FDA's recall record names no specific 510(k) clearance for this event. Product: HUDSON RCI mBrace, ET Tube Holder and 2 Point Head Strap, REF DYNJMBRC2; tracheal tube fixation device. Reason: Complaint defects have been noted regarding the adhesive elements of the device and the frame of the holder putting additional pressure on the patient's cheekbones and upper lip which may cause dermal injury depending on the device positioning..

Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Classification
Class II
Status
Ongoing
Initiated
2024-11-18
Firm location
Northfield, IL, United States

Product description

HUDSON RCI mBrace, ET Tube Holder and 2 Point Head Strap, REF DYNJMBRC2; tracheal tube fixation device

Reason for recall

Complaint defects have been noted regarding the adhesive elements of the device and the frame of the holder putting additional pressure on the patient's cheekbones and upper lip which may cause dermal injury depending on the device positioning.