Recall Z-0694-2018— Integra LifeSciences Corp.

Class II · Terminated

Class II FDA medical device recall by Integra LifeSciences Corp., initiated 2017-09-26, terminated 2018-08-14. It names one FDA 510(k) clearance: K161882. Product: Footswitch accessory to the CUSA Clarity Ultrasonic Tissue Ablation System; Catalogue No: C7000 Product Usage: The footswitch is a component of the CUSA Clarity Ultrasonic Surgical Aspirator System, which is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft tissue is desirable. The CUSA Clarity Ultrasonic Surgical Aspirator is indicated for use in: Neurosurgery, Plastic and Reconstructive surgery, Orthopedic Surgery, Gynecological Surgery and Thoracic Surgery and the following specific uses are cleared for use in the US only: Gastrointestinal and Affiliated Organ Surgery including removal of benign or malignant tumors or other unwanted tissue, including hepatic parenchyma, in open or laparoscopic procedures, hepatic resection, tumor resection, lobectomy or trisegmentectomy, or removal of tissue during liver allotransplantation and donor hepatectomy Urological surgery including removal of renal parenchyma during nephrectomy or partial nephrectomy General Surgery including removal of benign or malignant tumors or other unwanted soft tissue in open or minimally invasive general surgical procedures Laparoscopic Surgery including removal of hepatic parenchyma in laparoscopic hepatic resection, lobectomy or trisegmentectomy, in laparoscopic donor hepatectomy or laparoscopic cholecystectomy or laparoscopic pancreatic jejunostomy, or pancreatectomy, or laparoscopic appendectomy, laparoscopic colon resection or laparoscopic partial gastrectomy.. Reason: A faulty footswitch may result in unintended ultrasonic fragmentation activation. There are two scenarios when the failure mode may occur: 1) Ultrasonic fragmentation may be inadvertently activated at start-up or at the initial use once the footswitch is connected to the console and after the user primes the system, performs a handpiece test, and enters the Main Screen. 2) Ultrasonic fragmentation may remain activated during surgical use once the foot pedal is disengaged slowly..

Recalling firm
Integra LifeSciences Corp.
Classification
Class II
Status
Terminated
Initiated
2017-09-26
Terminated
2018-08-14
Firm location
Plainsboro, NJ, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Footswitch accessory to the CUSA Clarity Ultrasonic Tissue Ablation System; Catalogue No: C7000 Product Usage: The footswitch is a component of the CUSA Clarity Ultrasonic Surgical Aspirator System, which is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft tissue is desirable. The CUSA Clarity Ultrasonic Surgical Aspirator is indicated for use in: Neurosurgery, Plastic and Reconstructive surgery, Orthopedic Surgery, Gynecological Surgery and Thoracic Surgery and the following specific uses are cleared for use in the US only: Gastrointestinal and Affiliated Organ Surgery including removal of benign or malignant tumors or other unwanted tissue, including hepatic parenchyma, in open or laparoscopic procedures, hepatic resection, tumor resection, lobectomy or trisegmentectomy, or removal of tissue during liver allotransplantation and donor hepatectomy Urological surgery including removal of renal parenchyma during nephrectomy or partial nephrectomy General Surgery including removal of benign or malignant tumors or other unwanted soft tissue in open or minimally invasive general surgical procedures Laparoscopic Surgery including removal of hepatic parenchyma in laparoscopic hepatic resection, lobectomy or trisegmentectomy, in laparoscopic donor hepatectomy or laparoscopic cholecystectomy or laparoscopic pancreatic jejunostomy, or pancreatectomy, or laparoscopic appendectomy, laparoscopic colon resection or laparoscopic partial gastrectomy.

Reason for recall

A faulty footswitch may result in unintended ultrasonic fragmentation activation. There are two scenarios when the failure mode may occur: 1) Ultrasonic fragmentation may be inadvertently activated at start-up or at the initial use once the footswitch is connected to the console and after the user primes the system, performs a handpiece test, and enters the Main Screen. 2) Ultrasonic fragmentation may remain activated during surgical use once the foot pedal is disengaged slowly.

Data sourced from openFDA. This site is unofficial and independent of the FDA.