Recall Z-0687-2018— Cosman Medical, LLC
Class II · Terminated
Class II FDA medical device recall by Cosman Medical, LLC, initiated 2017-07-05, terminated 2018-09-14. It names one FDA 510(k) clearance: K050084. Product: Nitinol TC Reusable Electrodes (TCN), Model Numbers: 1)TCN-10, 2) TCNK-10-R*, 3)TCN-10-3M, 4) TCNK-15, 5) TCN-15, 6) TCNK-15-C, 7) TCN-15-3M, 8) TCNK-15-R*, 9) TCN-20, 10) TCNK-20, 11) TCN-20-3M*, 12) TCNK-20-C*, 13) TCN-5, 14) TCNK-20-R*, 15) TCN-5-3M, 16) TCNK-5, 17) TCNK-10, 18) TCNK-5-C, 19) TCNK-10-C & 20) TCNK-5-R*. The TCN devices are indicated for use in radiofrequency (RF) heat lesioning of peripheral nerve tissue only.. Reason: After multiple reprocessing cycles, the epoxy resin which holds the TCN Electrode in the hub can exhibit signs of damage. In some cases, this damage may result in the inability to fully remove blood and/or tissue residuals prior to cleaning and re-sterilizing the device..
- Recalling firm
- Cosman Medical, LLC
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-07-05
- Terminated
- 2018-09-14
- Firm location
- Burlington, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Nitinol TC Reusable Electrodes (TCN), Model Numbers: 1)TCN-10, 2) TCNK-10-R*, 3)TCN-10-3M, 4) TCNK-15, 5) TCN-15, 6) TCNK-15-C, 7) TCN-15-3M, 8) TCNK-15-R*, 9) TCN-20, 10) TCNK-20, 11) TCN-20-3M*, 12) TCNK-20-C*, 13) TCN-5, 14) TCNK-20-R*, 15) TCN-5-3M, 16) TCNK-5, 17) TCNK-10, 18) TCNK-5-C, 19) TCNK-10-C & 20) TCNK-5-R*. The TCN devices are indicated for use in radiofrequency (RF) heat lesioning of peripheral nerve tissue only.
Reason for recall
After multiple reprocessing cycles, the epoxy resin which holds the TCN Electrode in the hub can exhibit signs of damage. In some cases, this damage may result in the inability to fully remove blood and/or tissue residuals prior to cleaning and re-sterilizing the device.
Data sourced from openFDA. This site is unofficial and independent of the FDA.