Recall Z-0679-2025— Noah Medical

Class II · Ongoing

Class II FDA medical device recall by Noah Medical, initiated 2024-10-25. It names one FDA 510(k) clearance: K223144. Product: Galaxy bronchoscope (Product GALB-001) used with Instructions for use, Part: 10000752, part of the Noah Medical Galaxy System (GAL-001). Reason: Not all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive force may result in tissue injury/pneumothorax; instruments with long rigid distal sections may exhibit insertion resistance when bronchoscope bending section significantly articulated..

Recalling firm
Noah Medical
Classification
Class II
Status
Ongoing
Initiated
2024-10-25
Firm location
San Carlos, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Galaxy bronchoscope (Product GALB-001) used with Instructions for use, Part: 10000752, part of the Noah Medical Galaxy System (GAL-001)

Reason for recall

Not all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive force may result in tissue injury/pneumothorax; instruments with long rigid distal sections may exhibit insertion resistance when bronchoscope bending section significantly articulated.

Data sourced from openFDA. This site is unofficial and independent of the FDA.