Recall Z-0677-2026— B Braun Medical Inc

Class II · Ongoing

Class II FDA medical device recall by B Braun Medical Inc, initiated 2025-10-29. It names 2 FDA 510(k) clearances: K170595, K931377. Product: IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. 15DR 2SAFELINE 1CARESITE 1STOPCOCK 131IN; Catalog Number: 490612.. Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion)..

Recalling firm
B Braun Medical Inc
Classification
Class II
Status
Ongoing
Initiated
2025-10-29
Firm location
Bethlehem, PA, United States

510(k) clearances named in this recall

2 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. 15DR 2SAFELINE 1CARESITE 1STOPCOCK 131IN; Catalog Number: 490612.

Reason for recall

Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

Data sourced from openFDA. This site is unofficial and independent of the FDA.