Recall Z-0677-2025— Diasorin Inc.

Class II · Ongoing

Class II FDA medical device recall by Diasorin Inc., initiated 2024-11-07. It names one FDA 510(k) clearance: K213858. Product: DiaSorin LIAISON Q.S.E.T. Device Plus, REF 319060; intended for use in the preparation of human stool specimens. Reason: Customer complaints were received alleging leaking devices and insufficient buffer volume. Diasorin's internal investigation determined that 0.14% of the LIAISON Q.S.E.T. Device Plus in the lots identified have loose clear caps, which may allow buffer to leak from the tube. A stool sample prepared with less than the correct amount of buffer may result an incorrectly high patient result..

Recalling firm
Diasorin Inc.
Classification
Class II
Status
Ongoing
Initiated
2024-11-07
Firm location
Stillwater, MN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

DiaSorin LIAISON Q.S.E.T. Device Plus, REF 319060; intended for use in the preparation of human stool specimens

Reason for recall

Customer complaints were received alleging leaking devices and insufficient buffer volume. Diasorin's internal investigation determined that 0.14% of the LIAISON Q.S.E.T. Device Plus in the lots identified have loose clear caps, which may allow buffer to leak from the tube. A stool sample prepared with less than the correct amount of buffer may result an incorrectly high patient result.

Data sourced from openFDA. This site is unofficial and independent of the FDA.