Recall Z-0677-2024— Philips North America
Class II · Ongoing
Class II FDA medical device recall by Philips North America, initiated 2023-11-30. It names one FDA 510(k) clearance: K062298. Product: BrightView, Gamma Camera System, Product Code 882480.. Reason: While using Pre-Programmed Motion during an extrinsic quality assurance scan, a gap is created between the patient support and the detector, which may present a potential extremity entrapment hazard. Risk to patients may include fracture, body part loss of function/debilitation, muscle or ligament sprain or strain, laceration, crush injury, abrasion, or contusion..
- Recalling firm
- Philips North America
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-11-30
- Firm location
- Cambridge, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
BrightView, Gamma Camera System, Product Code 882480.
Reason for recall
While using Pre-Programmed Motion during an extrinsic quality assurance scan, a gap is created between the patient support and the detector, which may present a potential extremity entrapment hazard. Risk to patients may include fracture, body part loss of function/debilitation, muscle or ligament sprain or strain, laceration, crush injury, abrasion, or contusion.
Data sourced from openFDA. This site is unofficial and independent of the FDA.