Recall Z-0677-2022— Biomet, Inc.
Class II · Terminated
Class II FDA medical device recall by Biomet, Inc., initiated 2022-01-07, terminated 2023-04-13. FDA's recall record names no specific 510(k) clearance for this event. Product: Comprehensive Shoulder System Mini Humeral Stem, Porous Plasma, 15 MM, 83 MM Long Item Number: 113635. Reason: The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor..
- Recalling firm
- Biomet, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2022-01-07
- Terminated
- 2023-04-13
- Firm location
- Warsaw, IN, United States
Product description
Comprehensive Shoulder System Mini Humeral Stem, Porous Plasma, 15 MM, 83 MM Long Item Number: 113635
Reason for recall
The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.
Data sourced from openFDA. This site is unofficial and independent of the FDA.