Recall Z-0677-2016— GE Medical Systems, LLC
Class II · Terminated
Class II FDA medical device recall by GE Medical Systems, LLC, initiated 2015-12-15, terminated 2016-09-21. It names 2 FDA 510(k) clearances: K143530, K122445. Product: GE Healthcare, Aespire 7900, 1009-9012-000 & Aespire View, 1009-9212-000. The family of GE Datex-Ohmeda Aespire anesthesia systems with 7900 ventilator (Aespire 7900 and Aespire View) is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. The devices are intended for volume or pressure control ventilation.. Reason: The Drive Gas Check Valve of certain Anesthesia products could become stuck in a fixed open position, which could cause pressure to build in the mechanical ventilation cycle. If this issue is left unresolved, it could result in excessive or prolonged pressure in the patient breathing circuit during ventilation potentially resulting in barotrauma..
- Recalling firm
- GE Medical Systems, LLC
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2015-12-15
- Terminated
- 2016-09-21
- Firm location
- Waukesha, WI, United States
510(k) clearances named in this recall
2 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
GE Healthcare, Aespire 7900, 1009-9012-000 & Aespire View, 1009-9212-000. The family of GE Datex-Ohmeda Aespire anesthesia systems with 7900 ventilator (Aespire 7900 and Aespire View) is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. The devices are intended for volume or pressure control ventilation.
Reason for recall
The Drive Gas Check Valve of certain Anesthesia products could become stuck in a fixed open position, which could cause pressure to build in the mechanical ventilation cycle. If this issue is left unresolved, it could result in excessive or prolonged pressure in the patient breathing circuit during ventilation potentially resulting in barotrauma.