Recall Z-0673-2020— Leica Microsystems, Inc.

Class II · Ongoing

Class II FDA medical device recall by Leica Microsystems, Inc., initiated 2019-11-18. FDA's recall record names no specific 510(k) clearance for this event. Product: BOND Polymer Refine Detection Kit, Catalog No. DS9800. This detection system is for in vitro diagnostic use. Bond Polymer Refine Detection is a biotin-free, polymeric horseradish peroxidase (HRP)-linker antibody conjugate system for the detection of tissue-bound mouse and rabbit IgG and some mouse IgM primary antibodies. It is intended for staining sections of formalin-fixed, paraffin-embedded tissue on the BOND automated system.. Reason: BOND Polymer Refine Detection kits (DS9800) were held out of refrigerated storage longer than specified which may exhibit suboptimal staining..

Recalling firm
Leica Microsystems, Inc.
Classification
Class II
Status
Ongoing
Initiated
2019-11-18
Firm location
Buffalo Grove, IL, United States

Product description

BOND Polymer Refine Detection Kit, Catalog No. DS9800. This detection system is for in vitro diagnostic use. Bond Polymer Refine Detection is a biotin-free, polymeric horseradish peroxidase (HRP)-linker antibody conjugate system for the detection of tissue-bound mouse and rabbit IgG and some mouse IgM primary antibodies. It is intended for staining sections of formalin-fixed, paraffin-embedded tissue on the BOND automated system.

Reason for recall

BOND Polymer Refine Detection kits (DS9800) were held out of refrigerated storage longer than specified which may exhibit suboptimal staining.