Recall Z-0670-2026— B Braun Medical Inc

Class II · Ongoing

Class II FDA medical device recall by B Braun Medical Inc, initiated 2025-10-29. It names one FDA 510(k) clearance: K170595. Product: IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 15 DR W/2 SAFEDAY; Catalog Number: 352636. 2. IV ADMIN SET 15 DR W/2 SAFEDAY 84 IN.; Catalog Number: 352639. 3. IV ADMIN SET 15 DROP W/3 SAFEDAY; Catalog Number: 352640. 4. IV ADMIN 15D W/2 SAFEDAY & 0.2 FILTER; Catalog Number: 352643. 5. IV ADMIN SET, 105 IN.; Catalog Number: 477003. 6. IV ADMIN SET 15D, 4 CARESTIE LAD, 145 IN; Catalog Number: 477006. 7. IV ADMIN SET, 121 IN.; Catalog Number: 477008. 8. IV ADMIN SET 15D, CARESTIE LAD, 90 IN; Catalog Number: 477012. 9. IV ADMIN SET, 2 CARESTIE LAD, 15D, 117IN; Catalog Number: 477019. 10. IV ADMIN SET, 15D, 5CARESITE LAD, 136 IN; Catalog Number: 477021. 11. IV ADMIN SET 114 IN.; Catalog Number: 477022. 12. IV ADMIN SET 5 CARESITE LADS, 126 IN.; Catalog Number: 477023. 13. IV SET 15DR 3 SAFEDAY, 106 IN.; Catalog Number: 477053. 14. IV ADMIN SET, 15DR, CARESITE LAD, 121 IN; Catalog Number: 490513. 15. IV ADMIN SET, 15D, 120 IN.; Catalog Number: 490521. 16. IV ADMIN SET 15D 2 CARESITE LADS, 112IN.; Catalog Number: 490527. 17. IV ADMIN SET 15 DRP CARESITE SAFEPORT; Catalog Number: 490529. 18. IV ADMIN SET, 15D, 139 IN.; Catalog Number: 490535. 19. IV ADMIN SET 15D, CARESITE LAD, 122IN.; Catalog Number: 490543. 20. IV ADMIN SET 15D 3 CARESITE LADS, 150IN.; Catalog Number: 490546. 21. IV ADMIN SET, 125 IN.; Catalog Number: 490547. 22. IV ADMIN SET 15D, 5 CARESITE INJ, 150in.; Catalog Number: 490552. 23. IV ADMIN SET, 120IN.; Catalog Number: 490553. 24. IV ADMIN SETW/ B/C VLV 3 CARESITE 134IN.; Catalog Number: 490554. 25. IV ADMIN SET 15DR w/3 B/C VLV, 124 IN.; Catalog Number: 490555. 26. IV ADMIN SET 15D w/B/C VLV, 139 IN.; Catalog Number: 490558. 27. IV ADMINISTRATION SET, 117 IN.; Catalog Number: 490559. 28. IV ADMINISTRATION SET, 108 IN.; Catalog Number: 490560. 29. IV ADMINISTRATION SET, 134 IN.; Catalog Number: 490561. 30. IV ADMIN SET 15D w/B/C VLV, 140 IN.; Catalog Number: 490562. 31. IV ADMIN SET, 111 IN.; Catalog Number: 490563. 32. IV ADMIN SET 15D 2 CARESITE LADS, 122 IN; Catalog Number: 490569. 33. IV ADMIN SET 15D STOPCOCK, DUAL MANIFOLD; Catalog Number: 490573. 34. IV ADMIN SET 15D w/B/C VLV, 110 IN.; Catalog Number: 490574. 35. IV ADMIN 15D 5 CARESITE LADS STPCK 120in; Catalog Number: 490575. 36. IV ADMIN SET 15D 5 CARESITE LADS, 140 IN; Catalog Number: 490578. 37. IV ADMIN SET W/15D, 4 CARESITE LADS; Catalog Number: 490581. 38. IV ADMIN 15D CARESITE LADS, STPCK 134 IN; Catalog Number: 490582. 39. IV ADMIN SET, 15D, 108 IN.; Catalog Number: 490583. 40. IV ADMIN SET, 10 DROP, 130 IN.; Catalog Number: 490588. 41. IV ADMIN SET, 132 IN.; Catalog Number: 490589. 42. IV ADMIN SET, 15DR, 105 IN.; Catalog Number: 490600. 43. IV ADMIN SET,3 CARESITE,3 HF SPCK, 129IN; Catalog Number: 490607. 44. IV SET 15DR 3CARESITE 2STOPCOCK, 126IN; Catalog Number: 490608. 45. IV ADMIN SET 15D 2 CARESITE LADS, 138 IN; Catalog Number: 490615. 46. IV SET 15DR 2SAFEDAY 2SAFEPORT, 99 IN.; Catalog Number: 490618. 47. IV SET 60DR 2SAFEDAY 2SAFEPORT, 99 IN.; Catalog Number: 490619. 48. IV ADMIN SET 15 DRPS 4W STOPCOCK W/ C&S; Catalog Number: 490644. 49. IV ADMIN SET 15 DRPS ULTRAP. ZERO 90.6IN; Catalog Number: 490657. 50. IV ADMIN SET 15 DRPS STOPCOCK MALE 117IN; Catalog Number: 490658.. Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion)..

Recalling firm
B Braun Medical Inc
Classification
Class II
Status
Ongoing
Initiated
2025-10-29
Firm location
Bethlehem, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 15 DR W/2 SAFEDAY; Catalog Number: 352636. 2. IV ADMIN SET 15 DR W/2 SAFEDAY 84 IN.; Catalog Number: 352639. 3. IV ADMIN SET 15 DROP W/3 SAFEDAY; Catalog Number: 352640. 4. IV ADMIN 15D W/2 SAFEDAY & 0.2 FILTER; Catalog Number: 352643. 5. IV ADMIN SET, 105 IN.; Catalog Number: 477003. 6. IV ADMIN SET 15D, 4 CARESTIE LAD, 145 IN; Catalog Number: 477006. 7. IV ADMIN SET, 121 IN.; Catalog Number: 477008. 8. IV ADMIN SET 15D, CARESTIE LAD, 90 IN; Catalog Number: 477012. 9. IV ADMIN SET, 2 CARESTIE LAD, 15D, 117IN; Catalog Number: 477019. 10. IV ADMIN SET, 15D, 5CARESITE LAD, 136 IN; Catalog Number: 477021. 11. IV ADMIN SET 114 IN.; Catalog Number: 477022. 12. IV ADMIN SET 5 CARESITE LADS, 126 IN.; Catalog Number: 477023. 13. IV SET 15DR 3 SAFEDAY, 106 IN.; Catalog Number: 477053. 14. IV ADMIN SET, 15DR, CARESITE LAD, 121 IN; Catalog Number: 490513. 15. IV ADMIN SET, 15D, 120 IN.; Catalog Number: 490521. 16. IV ADMIN SET 15D 2 CARESITE LADS, 112IN.; Catalog Number: 490527. 17. IV ADMIN SET 15 DRP CARESITE SAFEPORT; Catalog Number: 490529. 18. IV ADMIN SET, 15D, 139 IN.; Catalog Number: 490535. 19. IV ADMIN SET 15D, CARESITE LAD, 122IN.; Catalog Number: 490543. 20. IV ADMIN SET 15D 3 CARESITE LADS, 150IN.; Catalog Number: 490546. 21. IV ADMIN SET, 125 IN.; Catalog Number: 490547. 22. IV ADMIN SET 15D, 5 CARESITE INJ, 150in.; Catalog Number: 490552. 23. IV ADMIN SET, 120IN.; Catalog Number: 490553. 24. IV ADMIN SETW/ B/C VLV 3 CARESITE 134IN.; Catalog Number: 490554. 25. IV ADMIN SET 15DR w/3 B/C VLV, 124 IN.; Catalog Number: 490555. 26. IV ADMIN SET 15D w/B/C VLV, 139 IN.; Catalog Number: 490558. 27. IV ADMINISTRATION SET, 117 IN.; Catalog Number: 490559. 28. IV ADMINISTRATION SET, 108 IN.; Catalog Number: 490560. 29. IV ADMINISTRATION SET, 134 IN.; Catalog Number: 490561. 30. IV ADMIN SET 15D w/B/C VLV, 140 IN.; Catalog Number: 490562. 31. IV ADMIN SET, 111 IN.; Catalog Number: 490563. 32. IV ADMIN SET 15D 2 CARESITE LADS, 122 IN; Catalog Number: 490569. 33. IV ADMIN SET 15D STOPCOCK, DUAL MANIFOLD; Catalog Number: 490573. 34. IV ADMIN SET 15D w/B/C VLV, 110 IN.; Catalog Number: 490574. 35. IV ADMIN 15D 5 CARESITE LADS STPCK 120in; Catalog Number: 490575. 36. IV ADMIN SET 15D 5 CARESITE LADS, 140 IN; Catalog Number: 490578. 37. IV ADMIN SET W/15D, 4 CARESITE LADS; Catalog Number: 490581. 38. IV ADMIN 15D CARESITE LADS, STPCK 134 IN; Catalog Number: 490582. 39. IV ADMIN SET, 15D, 108 IN.; Catalog Number: 490583. 40. IV ADMIN SET, 10 DROP, 130 IN.; Catalog Number: 490588. 41. IV ADMIN SET, 132 IN.; Catalog Number: 490589. 42. IV ADMIN SET, 15DR, 105 IN.; Catalog Number: 490600. 43. IV ADMIN SET,3 CARESITE,3 HF SPCK, 129IN; Catalog Number: 490607. 44. IV SET 15DR 3CARESITE 2STOPCOCK, 126IN; Catalog Number: 490608. 45. IV ADMIN SET 15D 2 CARESITE LADS, 138 IN; Catalog Number: 490615. 46. IV SET 15DR 2SAFEDAY 2SAFEPORT, 99 IN.; Catalog Number: 490618. 47. IV SET 60DR 2SAFEDAY 2SAFEPORT, 99 IN.; Catalog Number: 490619. 48. IV ADMIN SET 15 DRPS 4W STOPCOCK W/ C&S; Catalog Number: 490644. 49. IV ADMIN SET 15 DRPS ULTRAP. ZERO 90.6IN; Catalog Number: 490657. 50. IV ADMIN SET 15 DRPS STOPCOCK MALE 117IN; Catalog Number: 490658.

Reason for recall

Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).