Recall Z-0669-2020— Roche Diagnostics Operations, Inc.
Class II · Terminated
Class II FDA medical device recall by Roche Diagnostics Operations, Inc., initiated 2019-10-28, terminated 2022-06-28. It names one FDA 510(k) clearance: K093955. Product: Elecsys Anti-HAV IgM assay-for the in vitro qualitative detection of IgM antibodies to hepatitis A virus (anti-HAV IgM) in human serum and plasma Catalog number: 07026773190. Reason: Current software version of the Anti-HAV IgM APP-file (version 06.05-101) downloaded an erroneous Border Area setting, (1.0 1.00) instead of (0.90 1.10)..
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-10-28
- Terminated
- 2022-06-28
- Firm location
- Indianapolis, IN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Elecsys Anti-HAV IgM assay-for the in vitro qualitative detection of IgM antibodies to hepatitis A virus (anti-HAV IgM) in human serum and plasma Catalog number: 07026773190
Reason for recall
Current software version of the Anti-HAV IgM APP-file (version 06.05-101) downloaded an erroneous Border Area setting, (1.0 1.00) instead of (0.90 1.10).
Data sourced from openFDA. This site is unofficial and independent of the FDA.