Recall Z-0669-2013— Terumo Cardiovascular Systems Corporation
Class II · Terminated
Class II FDA medical device recall by Terumo Cardiovascular Systems Corporation, initiated 2013-01-07, terminated 2013-08-26. It names 3 FDA 510(k) clearances: K083301, K771499, K770429. Product: Sarns high-flow aortic arch cannula, 5.2 mm (16 Fr) OD with 3/8in connector, 7.5in (19 cm) long The Sarns" High Flow Aortic Arch Cannula and Sarns" Flexible Arterial Cannulae are indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. Reason: Flexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the potential to detach in the operative field, TCVS has decided to remove all affected lots that were manufactured with the affected connector part..
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-01-07
- Terminated
- 2013-08-26
- Firm location
- Ann Arbor, MI, United States
510(k) clearances named in this recall
3 on FDA's record- K083301 — X-COATED TERUMO AND SARNS DUAL STAGE VENOUS RETURN CATHETERS, SARNS MALLEABLE VENOUS AND DUAL STAGE RETURN CATHETERS
- K771499 — FLEXIBLE ARTERIAL CANNULAE
- K770429 — CANNULAE, AORTIC, ARCH, HIGH FLOW
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Sarns high-flow aortic arch cannula, 5.2 mm (16 Fr) OD with 3/8in connector, 7.5in (19 cm) long The Sarns" High Flow Aortic Arch Cannula and Sarns" Flexible Arterial Cannulae are indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery
Reason for recall
Flexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the potential to detach in the operative field, TCVS has decided to remove all affected lots that were manufactured with the affected connector part.
Data sourced from openFDA. This site is unofficial and independent of the FDA.