Recall Z-0665-2022— Becton Dickinson & Company

Class III · Ongoing

Class III FDA medical device recall by Becton Dickinson & Company, initiated 2022-01-06. It names one FDA 510(k) clearance: K991088. Product: BD Vacutainer Luer Lok Access Device - intended as a sterile, single use device designed to provide a connection to a catheter port equipped with a threaded female luer during blood or urine sample collection using evacuated BD Vacutainer Blood or Urine Collection Tubes Catalog Number: 364902, 36490200. Reason: May shed contaminant particles into the urine specimen that may be misidentified as Red Blood Cells when used in conjunction with the Sysmex UF-1000i or UX-2000 Analyzers and lead to false elevations in RBC counts.

Recalling firm
Becton Dickinson & Company
Classification
Class III
Status
Ongoing
Initiated
2022-01-06
Firm location
Franklin Lakes, NJ, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

BD Vacutainer Luer Lok Access Device - intended as a sterile, single use device designed to provide a connection to a catheter port equipped with a threaded female luer during blood or urine sample collection using evacuated BD Vacutainer Blood or Urine Collection Tubes Catalog Number: 364902, 36490200

Reason for recall

May shed contaminant particles into the urine specimen that may be misidentified as Red Blood Cells when used in conjunction with the Sysmex UF-1000i or UX-2000 Analyzers and lead to false elevations in RBC counts

Data sourced from openFDA. This site is unofficial and independent of the FDA.