Recall Z-0655-2014— The Anspach Effort, Inc.
Class II · Terminated
Class II FDA medical device recall by The Anspach Effort, Inc., initiated 2013-04-25, terminated 2015-07-20. It names one FDA 510(k) clearance: K965080. Product: ***REF XMAX***XMax Handpiece***Rx Only***Manufacturer: The Anspach Effort, Inc. Intended for cutting and shaping bone.. Reason: The firm Anspach Effort of Palm Beach Gardens, FL initiated a voluntary removal of the Pneumatic Drill System, including the motor hand-piece, Auto Lube foot control, without a pressure relief valve (PRV) due to the hand piece's outer hose can rupture without warning if the air flow is blocked during use...
- Recalling firm
- The Anspach Effort, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-04-25
- Terminated
- 2015-07-20
- Firm location
- Palm Beach Gardens, FL, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
***REF XMAX***XMax Handpiece***Rx Only***Manufacturer: The Anspach Effort, Inc. Intended for cutting and shaping bone.
Reason for recall
The firm Anspach Effort of Palm Beach Gardens, FL initiated a voluntary removal of the Pneumatic Drill System, including the motor hand-piece, Auto Lube foot control, without a pressure relief valve (PRV) due to the hand piece's outer hose can rupture without warning if the air flow is blocked during use..
Data sourced from openFDA. This site is unofficial and independent of the FDA.