Recall Z-0655-2013— Concentric Medical Inc
Class II · Terminated
Class II FDA medical device recall by Concentric Medical Inc, initiated 2012-12-20, terminated 2013-02-05. It names one FDA 510(k) clearance: K122581. Product: FlowGate Balloon Guide Catheter; Manufactured by Concentric Medical, Mountain View, CA. FlowGate" Balloon Guide Catheters are coaxial-lumen, braid-reinforced, variable stiffness catheters with a radiopaque marker on the distal end and a bifurcated luer hub on the proximal end. A balloon is flush mounted on the distal end. FlowGate" Balloon Guide Catheters are indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neurovascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for Retrieval devices.. Reason: Firm received complaints of resistance and possible collapse of the distal tip of the FlowGate Balloon Guide Catheters during use for aspiration..
- Recalling firm
- Concentric Medical Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2012-12-20
- Terminated
- 2013-02-05
- Firm location
- Mountain View, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
FlowGate Balloon Guide Catheter; Manufactured by Concentric Medical, Mountain View, CA. FlowGate" Balloon Guide Catheters are coaxial-lumen, braid-reinforced, variable stiffness catheters with a radiopaque marker on the distal end and a bifurcated luer hub on the proximal end. A balloon is flush mounted on the distal end. FlowGate" Balloon Guide Catheters are indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neurovascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for Retrieval devices.
Reason for recall
Firm received complaints of resistance and possible collapse of the distal tip of the FlowGate Balloon Guide Catheters during use for aspiration.