Recall Z-0645-2013— Horiba Instruments, Inc dba Horiba Medical

Class II · Terminated

Class II FDA medical device recall by Horiba Instruments, Inc dba Horiba Medical, initiated 2012-09-20, terminated 2013-03-12. FDA's recall record names no specific 510(k) clearance for this event. Product: ABX PENTRA ML, Software Versions: V8.0.x and Below; V.9.0.1 and V9.0.2. Reason: Horiba Medical is recalling ABX PENTRA ML because they have confirmed an issue with the ABX PENTRA ML data management and validation system. An incorrect patient result is transmitted to the Laboratory Information System. This result is from a previous run of the same patient that is stored in the ML database. This issue occurs in a very rare sequence of specific events. The affected software.

Recalling firm
Horiba Instruments, Inc dba Horiba Medical
Classification
Class II
Status
Terminated
Initiated
2012-09-20
Terminated
2013-03-12
Firm location
Irvine, CA, United States

Product description

ABX PENTRA ML, Software Versions: V8.0.x and Below; V.9.0.1 and V9.0.2

Reason for recall

Horiba Medical is recalling ABX PENTRA ML because they have confirmed an issue with the ABX PENTRA ML data management and validation system. An incorrect patient result is transmitted to the Laboratory Information System. This result is from a previous run of the same patient that is stored in the ML database. This issue occurs in a very rare sequence of specific events. The affected software