Recall Z-0641-2018— Allen Medical Systems
Class II · Terminated
Class II FDA medical device recall by Allen Medical Systems, initiated 2017-10-06, terminated 2019-04-03. FDA's recall record names no specific 510(k) clearance for this event. Product: Allen Medical Bow Frame Model A-70800. Surgical platform used to position patients for lumbar spine procedures.. Reason: The base of the frame has the potential to crack when the patient is initially transferred to the Bow Frame..
- Recalling firm
- Allen Medical Systems
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-10-06
- Terminated
- 2019-04-03
- Firm location
- Acton, MA, United States
Product description
Allen Medical Bow Frame Model A-70800. Surgical platform used to position patients for lumbar spine procedures.
Reason for recall
The base of the frame has the potential to crack when the patient is initially transferred to the Bow Frame.