Recall Z-0640-2026— B Braun Medical Inc
Class II · Ongoing
Class II FDA medical device recall by B Braun Medical Inc, initiated 2025-10-29. It names one FDA 510(k) clearance: K142036. Product: Infusomat utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. UNIV.15DROP PUMP SET W/3 SAFELINE LL; 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; Catalog Number: 362032. 2. UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; Catalog Number: 362033. 3. 60DROP METRISET PUMP SET, 3 SAFELINE; Priming volume: 19mL, length: 120in, SafeLine; Catalog Number: 362034. 4. Y TYPE BLOOD PUMP SET, SAFELINE INJ; Priming volume: 39mL, length: 120in, SafeLine¿; Catalog Number: 362043. 5. UNIV. 15 DROP PUMP SET W/ 0.2 FILTER; 15 drops/mL, priming volume: 26mL, length: 124in, SafeLine; Catalog Number: 362050. 6. SPACE PUMP SET 15 DROP W/2 SAFEDAY; SPACE PUMP SET 15 DROP W/2 SAFEDAY; Catalog Number: 362420. 7. UNIV. 15 DROP PUMP SET W/ 3 ULTRASITE LL; 15 drops/mL, priming volume: 21mL, length: 120in, ULTRASITE; Catalog Number: 362431. 8. SPACE PUMP IV SET, 120 IN.; 15 drops/mL, priming volume: 21mL, length: 118in, ULTRASITE; Catalog Number: 362432. 9. 60 DROP METRISET PUMP SET,3 SFLINE ASV; Priming volume: 21mL, length: 121in, SafeLine; Catalog Number: 363010. 10. Y-TYPE BLOOD PUMP SET, SAFELINE, ASV; Priming volume: 40mL, length: 123in, SafeLine; Catalog Number: 363019. 11. UNIV. 15 DROP PUMP SET W/3 SAFELINE, ASV; 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; Catalog Number: 363030. 12. UNIV. 15 DROP PUMP SET, 3 ULTRASITE, ASV; 15 drops/mL, priming volume: 22mL, length: 123in, ULTRASITE; Catalog Number: 363230. 13. Y-TYPE BLOOD PUMP SET, CARESITE, ASV; Priming volume: 41mL, length: 120in, CARESITE; Catalog Number: 363419. 14. SPACE PUMP IV SET, 2 CARESITES, ASV; 15 drops/mL, priming volume: 24mL, length: 120in, CARESITE; Catalog Number: 363420. 15. UNIV. 15 DROP PUMP SET, 2 CARESITE, ASV; 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; Catalog Number: 363421. 16. 60DROP METRISET PUMP, 3 CARESITES, ASV; Priming volume: 21mL, length: 130in, CARESITE; Catalog Number: 363422. 17. UNIV. 60 DROP PUMP SET, 2 CARESITES, ASV; 60 drops/mL, priming volume: 21mL, length: 118in, CARESITE; Catalog Number: 363423. 18. UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE ASV; UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE ASV; Catalog Number: 363424. 19. UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; 15 drops/mL, priming volume: 23mL, length: 123in, CARESITE; Catalog Number: 363430. 20. UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; Catalog Number: 363433. 21. SPACE PUMP IV SET 15D, 110 IN.; "SPACE PUMP IV SET 15D, 110 IN. CLOSE"; Catalog Number: 480254. 22. SPACE PUMP IV SET, 127 IN.; SPACE PUMP IV SET, 127 IN.; Catalog Number: 480255. 23. SPACE PUMP SET 15D, CARESITE LAD 122 IN. ; 15 drops/mL, priming volume: 23mL, length: 122in, CARESITE; Catalog Number: 480256. 24. 15 DROP PUMP SET 1 SAFEDAY 0.2 FILTER; 15 DROP PUMP SET 1 SAFEDAY 0.2 FILTER; Catalog Number: 480258. 25. SPACE PUMP IV SET 15D, 90 IN.; SPACE PUMP IV SET 15D, 90 IN.; Catalog Number: 480263. 26. UNIV. 15 DROP PUMP SET W 3 CARESITE L.L.; 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; Catalog Number: 490100. 27. 60 DROP METRISET PUMP SET, 3 CARESITES; Priming volume: 19mL, length: 120in, CARESITE; Catalog Number: 490101.. Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion)..
- Recalling firm
- B Braun Medical Inc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-10-29
- Firm location
- Bethlehem, PA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Infusomat utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. UNIV.15DROP PUMP SET W/3 SAFELINE LL; 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; Catalog Number: 362032. 2. UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE; Catalog Number: 362033. 3. 60DROP METRISET PUMP SET, 3 SAFELINE; Priming volume: 19mL, length: 120in, SafeLine; Catalog Number: 362034. 4. Y TYPE BLOOD PUMP SET, SAFELINE INJ; Priming volume: 39mL, length: 120in, SafeLine¿; Catalog Number: 362043. 5. UNIV. 15 DROP PUMP SET W/ 0.2 FILTER; 15 drops/mL, priming volume: 26mL, length: 124in, SafeLine; Catalog Number: 362050. 6. SPACE PUMP SET 15 DROP W/2 SAFEDAY; SPACE PUMP SET 15 DROP W/2 SAFEDAY; Catalog Number: 362420. 7. UNIV. 15 DROP PUMP SET W/ 3 ULTRASITE LL; 15 drops/mL, priming volume: 21mL, length: 120in, ULTRASITE; Catalog Number: 362431. 8. SPACE PUMP IV SET, 120 IN.; 15 drops/mL, priming volume: 21mL, length: 118in, ULTRASITE; Catalog Number: 362432. 9. 60 DROP METRISET PUMP SET,3 SFLINE ASV; Priming volume: 21mL, length: 121in, SafeLine; Catalog Number: 363010. 10. Y-TYPE BLOOD PUMP SET, SAFELINE, ASV; Priming volume: 40mL, length: 123in, SafeLine; Catalog Number: 363019. 11. UNIV. 15 DROP PUMP SET W/3 SAFELINE, ASV; 15 drops/mL, priming volume: 22mL, length: 119in, SafeLine; Catalog Number: 363030. 12. UNIV. 15 DROP PUMP SET, 3 ULTRASITE, ASV; 15 drops/mL, priming volume: 22mL, length: 123in, ULTRASITE; Catalog Number: 363230. 13. Y-TYPE BLOOD PUMP SET, CARESITE, ASV; Priming volume: 41mL, length: 120in, CARESITE; Catalog Number: 363419. 14. SPACE PUMP IV SET, 2 CARESITES, ASV; 15 drops/mL, priming volume: 24mL, length: 120in, CARESITE; Catalog Number: 363420. 15. UNIV. 15 DROP PUMP SET, 2 CARESITE, ASV; 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; Catalog Number: 363421. 16. 60DROP METRISET PUMP, 3 CARESITES, ASV; Priming volume: 21mL, length: 130in, CARESITE; Catalog Number: 363422. 17. UNIV. 60 DROP PUMP SET, 2 CARESITES, ASV; 60 drops/mL, priming volume: 21mL, length: 118in, CARESITE; Catalog Number: 363423. 18. UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE ASV; UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE ASV; Catalog Number: 363424. 19. UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; 15 drops/mL, priming volume: 23mL, length: 123in, CARESITE; Catalog Number: 363430. 20. UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV; Catalog Number: 363433. 21. SPACE PUMP IV SET 15D, 110 IN.; "SPACE PUMP IV SET 15D, 110 IN. CLOSE"; Catalog Number: 480254. 22. SPACE PUMP IV SET, 127 IN.; SPACE PUMP IV SET, 127 IN.; Catalog Number: 480255. 23. SPACE PUMP SET 15D, CARESITE LAD 122 IN. ; 15 drops/mL, priming volume: 23mL, length: 122in, CARESITE; Catalog Number: 480256. 24. 15 DROP PUMP SET 1 SAFEDAY 0.2 FILTER; 15 DROP PUMP SET 1 SAFEDAY 0.2 FILTER; Catalog Number: 480258. 25. SPACE PUMP IV SET 15D, 90 IN.; SPACE PUMP IV SET 15D, 90 IN.; Catalog Number: 480263. 26. UNIV. 15 DROP PUMP SET W 3 CARESITE L.L.; 15 drops/mL, priming volume: 22mL, length: 120in, CARESITE; Catalog Number: 490100. 27. 60 DROP METRISET PUMP SET, 3 CARESITES; Priming volume: 19mL, length: 120in, CARESITE; Catalog Number: 490101.
Reason for recall
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).