Recall Z-0638-2026— B Braun Medical Inc

Class II · Ongoing

Class II FDA medical device recall by B Braun Medical Inc, initiated 2025-10-29. FDA's recall record names no specific 510(k) clearance for this event. Product: Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. Caresite Extension Set_Export; Catalog Number: 470183.. Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion)..

Recalling firm
B Braun Medical Inc
Classification
Class II
Status
Ongoing
Initiated
2025-10-29
Firm location
Bethlehem, PA, United States

Product description

Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. Caresite Extension Set_Export; Catalog Number: 470183.

Reason for recall

Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

Data sourced from openFDA. This site is unofficial and independent of the FDA.