Recall Z-0634-2022— Synthes (USA) Products LLC

Class II · Ongoing

Class II FDA medical device recall by Synthes (USA) Products LLC, initiated 2021-12-22. FDA's recall record names no specific 510(k) clearance for this event. Product: Universal Chuck-The Universal Chuck is an Instrument Handle provided as an alternative instrument available for use in the TN-ADVANCED" Tibial Nailing System (TNA), TFN-ADVANCED" Proximal Femoral Nailing System (TFNA), RFN-ADVANCED" Retrograde Femoral Nailing System (RFNA), FRN-ADVANCED" Femoral Recon Nailing System (FRNA), and Flexible Monobloc Reamers. Instrument Handles are intended for holding instruments for manipulation during orthopedic surgery. Product No.: 03.043.001. Reason: Top cap may loosen and detach if the Universal Chuck becomes jammed and the user attempts to manually free the device, internal ball bearings may fall out onto the surgical field and cause surgical delay or, and if not recovered, adverse tissue reaction and infection..

Recalling firm
Synthes (USA) Products LLC
Classification
Class II
Status
Ongoing
Initiated
2021-12-22
Firm location
West Chester, PA, United States

Product description

Universal Chuck-The Universal Chuck is an Instrument Handle provided as an alternative instrument available for use in the TN-ADVANCED" Tibial Nailing System (TNA), TFN-ADVANCED" Proximal Femoral Nailing System (TFNA), RFN-ADVANCED" Retrograde Femoral Nailing System (RFNA), FRN-ADVANCED" Femoral Recon Nailing System (FRNA), and Flexible Monobloc Reamers. Instrument Handles are intended for holding instruments for manipulation during orthopedic surgery. Product No.: 03.043.001

Reason for recall

Top cap may loosen and detach if the Universal Chuck becomes jammed and the user attempts to manually free the device, internal ball bearings may fall out onto the surgical field and cause surgical delay or, and if not recovered, adverse tissue reaction and infection.

Data sourced from openFDA. This site is unofficial and independent of the FDA.