Recall Z-0634-2016— Perkinelmer

Class II · Terminated

Class II FDA medical device recall by Perkinelmer, initiated 2015-12-08, terminated 2017-08-04. FDA's recall record names no specific 510(k) clearance for this event. Product: Perkin Elmer ClWIZARD2 5-detector, 550 samples Product Code: 2470-0050, 3470-0050. The Wizard2 gamma counter is intended to detect and count gamma radiation emitted by clinical samples.. Reason: The content of the second #023 barcode ID label in the Barcode ID Label Binder is incorrect. The second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. If measurement protocols have been programmed for both ID #023 and #024, the protocol associated with barcode ID #024 is executed. If the error is undetected, the Gamma Counter may produce erroneous results..

Recalling firm
Perkinelmer
Classification
Class II
Status
Terminated
Initiated
2015-12-08
Terminated
2017-08-04
Firm location
Waltham, MA, United States

Product description

Perkin Elmer ClWIZARD2 5-detector, 550 samples Product Code: 2470-0050, 3470-0050. The Wizard2 gamma counter is intended to detect and count gamma radiation emitted by clinical samples.

Reason for recall

The content of the second #023 barcode ID label in the Barcode ID Label Binder is incorrect. The second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. If measurement protocols have been programmed for both ID #023 and #024, the protocol associated with barcode ID #024 is executed. If the error is undetected, the Gamma Counter may produce erroneous results.

Data sourced from openFDA. This site is unofficial and independent of the FDA.