Recall Z-0632-2016— Perkinelmer

Class II · Terminated

Class II FDA medical device recall by Perkinelmer, initiated 2015-12-08, terminated 2017-08-04. FDA's recall record names no specific 510(k) clearance for this event. Product: Perkin Elmer WIZARD2 1-detector, 550 samples Product Code: 2470-0010 The Wizard gamma counter is intended to detect and count gamma radiation emitted by clinical samples. Reason: The content of the second #023 barcode ID label in the Barcode ID Label Binder is incorrect. The second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. If measurement protocols have been programmed for both ID #023 and #024, the protocol associated with barcode ID #024 is executed. If the error is undetected, the Gamma Counter may produce erroneous results..

Recalling firm
Perkinelmer
Classification
Class II
Status
Terminated
Initiated
2015-12-08
Terminated
2017-08-04
Firm location
Waltham, MA, United States

Product description

Perkin Elmer WIZARD2 1-detector, 550 samples Product Code: 2470-0010 The Wizard gamma counter is intended to detect and count gamma radiation emitted by clinical samples

Reason for recall

The content of the second #023 barcode ID label in the Barcode ID Label Binder is incorrect. The second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. If measurement protocols have been programmed for both ID #023 and #024, the protocol associated with barcode ID #024 is executed. If the error is undetected, the Gamma Counter may produce erroneous results.