Recall Z-0621-2025— Becton Dickinson & Co.
Class II · Ongoing
Class II FDA medical device recall by Becton Dickinson & Co., initiated 2024-11-06. FDA's recall record names no specific 510(k) clearance for this event. Product: BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) are used for semi-quantitative in vitro susceptibility testing by the agar disc diffusion test procedure of common, rapidly growing and certain fastidious bacterial pathogens. Catalog Number: 231263. Reason: BD identified through potency testing as part of a stability test request to monitor Ampicillin AM-2 due to decrease in potency results of 65% at 18 months, may result in falsely resistant result for ampicillin susceptibility.
- Recalling firm
- Becton Dickinson & Co.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2024-11-06
- Firm location
- Sparks, MD, United States
Product description
BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) are used for semi-quantitative in vitro susceptibility testing by the agar disc diffusion test procedure of common, rapidly growing and certain fastidious bacterial pathogens. Catalog Number: 231263
Reason for recall
BD identified through potency testing as part of a stability test request to monitor Ampicillin AM-2 due to decrease in potency results of 65% at 18 months, may result in falsely resistant result for ampicillin susceptibility