Recall Z-0621-2014— Integra LifeSciences Corp.
Class II · Terminated
Class II FDA medical device recall by Integra LifeSciences Corp., initiated 2013-11-13, terminated 2015-02-13. FDA's recall record names no specific 510(k) clearance for this event. Product: Integra Small Rickham -Style Reservoir Rx Only Sterile for Single Use Only Integra NeuroSciences 311 Enterprise Drive Plainsboro, NJ 08536 USA Made in the USA NL 850-1121 Provides access to the lateral cerebral ventricles via hypodermic puncture. It is useful in obtaining CSF (cerebral spinal fluid_ samples for cytological and chemical studies, for monitoring ventricular fluid pressure and for ventricular drainage. The reservoir provides easy access to the lateral ventricles and to cystic tumors for the injections of chemotherapeutic agents and/or radio-isotopes. The rickham reservoir may be utilized in hydrocephalic patients as a component in systems designed to shunt CSF from the lateral ventricles into either the right atrium of the heart of the peritoneum.. Reason: Integra became aware through an adverse complaint trend that Rickham Style Reservoirs with bent or broken connectors that were shipped in sleeve-type package were found prior to use but after the sterile package was open.
- Recalling firm
- Integra LifeSciences Corp.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-11-13
- Terminated
- 2015-02-13
- Firm location
- Plainsboro, NJ, United States
Product description
Integra Small Rickham -Style Reservoir Rx Only Sterile for Single Use Only Integra NeuroSciences 311 Enterprise Drive Plainsboro, NJ 08536 USA Made in the USA NL 850-1121 Provides access to the lateral cerebral ventricles via hypodermic puncture. It is useful in obtaining CSF (cerebral spinal fluid_ samples for cytological and chemical studies, for monitoring ventricular fluid pressure and for ventricular drainage. The reservoir provides easy access to the lateral ventricles and to cystic tumors for the injections of chemotherapeutic agents and/or radio-isotopes. The rickham reservoir may be utilized in hydrocephalic patients as a component in systems designed to shunt CSF from the lateral ventricles into either the right atrium of the heart of the peritoneum.
Reason for recall
Integra became aware through an adverse complaint trend that Rickham Style Reservoirs with bent or broken connectors that were shipped in sleeve-type package were found prior to use but after the sterile package was open
Data sourced from openFDA. This site is unofficial and independent of the FDA.