Recall Z-0617-2024— Steris Corporation

Class II · Ongoing

Class II FDA medical device recall by Steris Corporation, initiated 2023-11-13. It names 3 FDA 510(k) clearances: K172677, K102996, K082988. Product: Hookup Accessories used with MEDIVATORS ADVANTAGE PLUS and ADVANTAGE PLUS Pass Thru Automated Endoscope Reprocessors Model # Model Description 2-8-001 HU ZERO CHANNEL NO LT 2-8-010 ASSY, HOOK-UP, OLYMPUS 2-8-014 HOOKUP, OPTIM LT 2-8-030 ASSY, HOOKUP, PENTAX-0K 2-8-040 ASSY, HKUP STORZ 0K 2-8-070 ASSY, HU VISION SCIENCE 2-8-112 ASSY, HOOKUP OLMPUS 1K 2-8-120 ASSY, HOOKUP FUJI 1K 2-8-210 ASSY, HOOK-UP OLYMPUS 2K 2-8-213 ASSY, HOOKUP, OLYMPUS-2K 2-8-218 ASSY HOOKUP OLY 2K 2-8-236 ASSY, HOOKUP FUJINON 2K 2-8-244 ASSY, HOOKUP, STORZ 2K 2-8-245 ASSY, HOOK-UP, STORZ 2K 2-8-246 ASSY, HKUP, PENTAX-2K 2-8-260 ASSY, HOOK-UP WOLF 2-K 2-8-280 ASSY HOOKUP 2-8-312 ASSY, HOOK-UP OLYMPUS-3K 2-8-330 ASSY, HOOK-UP, PENTAX-3K 2-8-332 ASSEMBLY, HOOK-UP, PENTAX 2-8-334 ASSEMBLY, HOOK-UP, PENTAX 2-8-335 ASSY, HOOKUP, OLYMPUS, 3K 2-8-336 ASSY,HKUP FUJI 70 BRONC 2-8-338 ASSY, HU, FUJI BRONCH U/S 2-8-421 ASSY, HOOKUP, FUJIFILM 2-8-428 ASSY, ED-580XX 2-8-431 ASSY, HOOKUP, PENTAX 2-8-432 ASSY, HOOK-UP, PENTAX 2-8-460 HOOKUP, ASSY WOLF 4K 2-8-511 ASSY, HOOK-UP, OLY 2-8-513 ASSY HOOKUP OLY 4K ESOPHA 2-8-522 ASSY, FUJINON 4K-JET 2-8-523 ASSY, HKUP, FUJI 4K SER 5 2-8-526 ASSY, HOOKUP FUJI 4K+JET 2-8-527 ASSY, FUJI 700 NO FWJ 2-8-528 HOOK, EI-580BT 2-8-534 ASSY, HOOK-UP, PEN-4K+JET 2-8-537 ASSY, HKUP, PTX90 4K+JET 2-8-539 HOOK, PTX I20 FWJ 2-8-540 ASSY-HOOK-UP,STORZ,5K 2-8-611 ASSY, HOOK-UP, OLY 2-8-612 ASSY, HOOKUP, PENTAX 2-8-614 ASM HU OLY TGF-UC180J 2-8-617 ASSY, HKUP, OLY 2 BIOPSY 2-8-623 ASSY,HKUP,FUJI-5 4K+JET 2-8-625 ASSY, HOOKUP, EG-580UT 2-8-626 ASSY, HOOKUP, EG-580UR 2-8-627 ASSY, HKUP, FUJI 700 SER 2-8-634 ASSY, HOOKUP PENTAX DUAL 2-8-639 HOOK, EG3X-J10 2-8-640 ASSY,HOOK-UP STORZ 6K 2-8-711 ASSY, HOOKUP, OLYMPUS 5K 2-8-718 ASSY, HOOKUP, OLY 160/180 2-8-740 ASSY, HOOK-UP, STORZ 2-8-811 ASSY HKUP OLY-ECHO+LIFT 2-8-901 KIT TEE/TOE/RUS HU ADV+ 5-8-010 ASSY, HOOKUP DBL LT 5-8-030 ASSY, HOOKUP 5-8-040 ASSY HOOKUP STORZ 0K DBL 5-8-050 ASSY, HOOKUP, DILATOR 5-8-112 HOOKUP 2-SCOPE 1-K 5-8-190 HOOK DBL, IS, ION 5-8-210 ASSY, HOOKUP 5-8-245 HOOKUP KARLSTORZ-2K 2-8-001HAN HU ZERO CHANNEL NO LT 2-8-002HAN ADVANTAGE TEST BLOCK 2-8-010HAN ASSEMBLY, HOOK-UP,OLYMPUS 2-8-030HAN ASSY,HOOKUP,PENTAX-OK 2-8-112HAN ASSY, HOOK-UP OLYMPUS 1K 2-8-160HAN ASSY, ADV HOOKUP, WOLF 2-8-210HAN ASSY, HOOK-UP OLYMPUS 2K 2-8-213HAN ASSY, HOOKUP, OLYMPUS-2K 2-8-218HAN ASSY HOOKUP OLY 2K 2-8-245HAN ASSY, HOOKUP, STORZ-2K 2-8-330HAN ASSY, HOOK-UP, PENTAX-3K 2-8-334HAN ASSEMBLY, HOOK-UP, PENTAX 2-8-335HAN ASSY,HOOKUP,OLYMPUS 2-8-336HAN ASSY,HKUP FUJI 70 BRONC 2-8-338HAN ASSY, HU, FUJI BRONCH U/S 2-8-421HAN ASSY, HOOKUP, FUJINON 2-8-428CAS HOOK, CAS, ED-580XX 2-8-428HAN HOOK, HAN, ED-580XX 2-8-431CAS ASSY, CASSETTE, PENTAX 2-8-431HAN ASSY, HOOK-UP, PENTAX 2-8-511HAN ASSY, HOOK-UP, OLY 2-8-526HAN ASSY, HOOKUP FUJI 4K+JET 2-8-536HAN HOOKUP,PENTAX,70ULT, 5K 2-8-537CAS ASSY,CASS,PTX 90 4K+JET 2-8-537HAN ASSY HKUP,PTX90 4K+JET 2-8-611HAN ASSY, HOOK-UP, OLY 2-8-626CAS ASSY, CASS, FUJINON 580UR 2-8-627HAN ASSY, HAN, FUJI 700 SER 2-8-634CAS ASSY,CASS PENTAX DUAL CH 2-8-634HAN ASSY,HOOKUP PENTAX DUAL 2-8-640HAN ASSY, HOOK-UP STORZ 6K 2-8-711HAN ASSY, HOOKUP, OLYMPUS-5K 2-8-718HAN ASSY,HOOK-UP, OLY 160/180 2-8-811HAN ASSY HKUP OLY-ECHO W/LIFT 2-8-902H HU BLOCK ULTRASOUND PROBE 5-8-030HAN ASSY, HOOKUP 5-8-050HAN ASSY,HOOKUP, DILATOR 5-8-112HAN ASSY, HOOKUP 2-SCOPE 1-K 5-8-244HAN ASSY, HKUP 2K DBL. Reason: Identification label on affected hookups could become illegible over time due to the label's ink coming off. This may potentially prevent users from being able to effectively identify the hookup and may ultimately result in postponement of patient procedures..

Recalling firm
Steris Corporation
Classification
Class II
Status
Ongoing
Initiated
2023-11-13
Firm location
Mentor, OH, United States

510(k) clearances named in this recall

3 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Hookup Accessories used with MEDIVATORS ADVANTAGE PLUS and ADVANTAGE PLUS Pass Thru Automated Endoscope Reprocessors Model # Model Description 2-8-001 HU ZERO CHANNEL NO LT 2-8-010 ASSY, HOOK-UP, OLYMPUS 2-8-014 HOOKUP, OPTIM LT 2-8-030 ASSY, HOOKUP, PENTAX-0K 2-8-040 ASSY, HKUP STORZ 0K 2-8-070 ASSY, HU VISION SCIENCE 2-8-112 ASSY, HOOKUP OLMPUS 1K 2-8-120 ASSY, HOOKUP FUJI 1K 2-8-210 ASSY, HOOK-UP OLYMPUS 2K 2-8-213 ASSY, HOOKUP, OLYMPUS-2K 2-8-218 ASSY HOOKUP OLY 2K 2-8-236 ASSY, HOOKUP FUJINON 2K 2-8-244 ASSY, HOOKUP, STORZ 2K 2-8-245 ASSY, HOOK-UP, STORZ 2K 2-8-246 ASSY, HKUP, PENTAX-2K 2-8-260 ASSY, HOOK-UP WOLF 2-K 2-8-280 ASSY HOOKUP 2-8-312 ASSY, HOOK-UP OLYMPUS-3K 2-8-330 ASSY, HOOK-UP, PENTAX-3K 2-8-332 ASSEMBLY, HOOK-UP, PENTAX 2-8-334 ASSEMBLY, HOOK-UP, PENTAX 2-8-335 ASSY, HOOKUP, OLYMPUS, 3K 2-8-336 ASSY,HKUP FUJI 70 BRONC 2-8-338 ASSY, HU, FUJI BRONCH U/S 2-8-421 ASSY, HOOKUP, FUJIFILM 2-8-428 ASSY, ED-580XX 2-8-431 ASSY, HOOKUP, PENTAX 2-8-432 ASSY, HOOK-UP, PENTAX 2-8-460 HOOKUP, ASSY WOLF 4K 2-8-511 ASSY, HOOK-UP, OLY 2-8-513 ASSY HOOKUP OLY 4K ESOPHA 2-8-522 ASSY, FUJINON 4K-JET 2-8-523 ASSY, HKUP, FUJI 4K SER 5 2-8-526 ASSY, HOOKUP FUJI 4K+JET 2-8-527 ASSY, FUJI 700 NO FWJ 2-8-528 HOOK, EI-580BT 2-8-534 ASSY, HOOK-UP, PEN-4K+JET 2-8-537 ASSY, HKUP, PTX90 4K+JET 2-8-539 HOOK, PTX I20 FWJ 2-8-540 ASSY-HOOK-UP,STORZ,5K 2-8-611 ASSY, HOOK-UP, OLY 2-8-612 ASSY, HOOKUP, PENTAX 2-8-614 ASM HU OLY TGF-UC180J 2-8-617 ASSY, HKUP, OLY 2 BIOPSY 2-8-623 ASSY,HKUP,FUJI-5 4K+JET 2-8-625 ASSY, HOOKUP, EG-580UT 2-8-626 ASSY, HOOKUP, EG-580UR 2-8-627 ASSY, HKUP, FUJI 700 SER 2-8-634 ASSY, HOOKUP PENTAX DUAL 2-8-639 HOOK, EG3X-J10 2-8-640 ASSY,HOOK-UP STORZ 6K 2-8-711 ASSY, HOOKUP, OLYMPUS 5K 2-8-718 ASSY, HOOKUP, OLY 160/180 2-8-740 ASSY, HOOK-UP, STORZ 2-8-811 ASSY HKUP OLY-ECHO+LIFT 2-8-901 KIT TEE/TOE/RUS HU ADV+ 5-8-010 ASSY, HOOKUP DBL LT 5-8-030 ASSY, HOOKUP 5-8-040 ASSY HOOKUP STORZ 0K DBL 5-8-050 ASSY, HOOKUP, DILATOR 5-8-112 HOOKUP 2-SCOPE 1-K 5-8-190 HOOK DBL, IS, ION 5-8-210 ASSY, HOOKUP 5-8-245 HOOKUP KARLSTORZ-2K 2-8-001HAN HU ZERO CHANNEL NO LT 2-8-002HAN ADVANTAGE TEST BLOCK 2-8-010HAN ASSEMBLY, HOOK-UP,OLYMPUS 2-8-030HAN ASSY,HOOKUP,PENTAX-OK 2-8-112HAN ASSY, HOOK-UP OLYMPUS 1K 2-8-160HAN ASSY, ADV HOOKUP, WOLF 2-8-210HAN ASSY, HOOK-UP OLYMPUS 2K 2-8-213HAN ASSY, HOOKUP, OLYMPUS-2K 2-8-218HAN ASSY HOOKUP OLY 2K 2-8-245HAN ASSY, HOOKUP, STORZ-2K 2-8-330HAN ASSY, HOOK-UP, PENTAX-3K 2-8-334HAN ASSEMBLY, HOOK-UP, PENTAX 2-8-335HAN ASSY,HOOKUP,OLYMPUS 2-8-336HAN ASSY,HKUP FUJI 70 BRONC 2-8-338HAN ASSY, HU, FUJI BRONCH U/S 2-8-421HAN ASSY, HOOKUP, FUJINON 2-8-428CAS HOOK, CAS, ED-580XX 2-8-428HAN HOOK, HAN, ED-580XX 2-8-431CAS ASSY, CASSETTE, PENTAX 2-8-431HAN ASSY, HOOK-UP, PENTAX 2-8-511HAN ASSY, HOOK-UP, OLY 2-8-526HAN ASSY, HOOKUP FUJI 4K+JET 2-8-536HAN HOOKUP,PENTAX,70ULT, 5K 2-8-537CAS ASSY,CASS,PTX 90 4K+JET 2-8-537HAN ASSY HKUP,PTX90 4K+JET 2-8-611HAN ASSY, HOOK-UP, OLY 2-8-626CAS ASSY, CASS, FUJINON 580UR 2-8-627HAN ASSY, HAN, FUJI 700 SER 2-8-634CAS ASSY,CASS PENTAX DUAL CH 2-8-634HAN ASSY,HOOKUP PENTAX DUAL 2-8-640HAN ASSY, HOOK-UP STORZ 6K 2-8-711HAN ASSY, HOOKUP, OLYMPUS-5K 2-8-718HAN ASSY,HOOK-UP, OLY 160/180 2-8-811HAN ASSY HKUP OLY-ECHO W/LIFT 2-8-902H HU BLOCK ULTRASOUND PROBE 5-8-030HAN ASSY, HOOKUP 5-8-050HAN ASSY,HOOKUP, DILATOR 5-8-112HAN ASSY, HOOKUP 2-SCOPE 1-K 5-8-244HAN ASSY, HKUP 2K DBL

Reason for recall

Identification label on affected hookups could become illegible over time due to the label's ink coming off. This may potentially prevent users from being able to effectively identify the hookup and may ultimately result in postponement of patient procedures.