Recall Z-0615-2024— Medtronic Neuromodulation

Class II · Ongoing

Class II FDA medical device recall by Medtronic Neuromodulation, initiated 2023-11-13. FDA's recall record names no specific 510(k) clearance for this event. Product: Medtronic SynchroMed II, Model 8637-40, Programmable pump. Reason: Medtronic recently identified that if the SynchroMed II pump switches into telemetry mode due to electromagnetic interference (EMI) from an MRI scan, while the pump is sounding an alarm, the pump will not resume drug delivery after leaving the MRI magnetic field, which is inconsistent with the current labeling..

Recalling firm
Medtronic Neuromodulation
Classification
Class II
Status
Ongoing
Initiated
2023-11-13
Firm location
Minneapolis, MN, United States

Product description

Medtronic SynchroMed II, Model 8637-40, Programmable pump

Reason for recall

Medtronic recently identified that if the SynchroMed II pump switches into telemetry mode due to electromagnetic interference (EMI) from an MRI scan, while the pump is sounding an alarm, the pump will not resume drug delivery after leaving the MRI magnetic field, which is inconsistent with the current labeling.

Data sourced from openFDA. This site is unofficial and independent of the FDA.