Recall Z-0612-2013— Beckman Coulter Inc.
Class II · Terminated
Class II FDA medical device recall by Beckman Coulter Inc., initiated 2012-11-01, terminated 2013-12-23. It names 3 FDA 510(k) clearances: K060256, K023049, K922823. Product: Access Immunoassay System, Access 2 Immunoassay System, Synchron LXi 725 Clinical System, UniCel DxC 600i Synchron Access Clinical System Part Numbers: Access 81600; Access 2 81600N, 386220, A69186, A25656, A12900; Access 2 (LXi) A15642; Access 2 Section, DxC 600i A25640. The Access Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids. The Synchron LXi and the UniCel DxC 600i are Access systems integrated with a chemistry analyzer.. Reason: It has been determined that, in some cases, pulley drive components have broken parts or have become loose, mispositioned, or disconnected from their drive shafts. When these defects occur, the pulley may not turn at all, or may turn at an incorrect speed, resulting in inadequate washing or mixing of the contents of the reaction vessel..
- Recalling firm
- Beckman Coulter Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2012-11-01
- Terminated
- 2013-12-23
- Firm location
- Brea, CA, United States
510(k) clearances named in this recall
3 on FDA's record- K060256 — UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM
- K023049 — SYNCHRON LX I 725 CLINICAL SYSTEM
- K922823 — ACCESS THEOPHYLLINE IMMUNOASSAY REAGENTS
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Access Immunoassay System, Access 2 Immunoassay System, Synchron LXi 725 Clinical System, UniCel DxC 600i Synchron Access Clinical System Part Numbers: Access 81600; Access 2 81600N, 386220, A69186, A25656, A12900; Access 2 (LXi) A15642; Access 2 Section, DxC 600i A25640. The Access Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids. The Synchron LXi and the UniCel DxC 600i are Access systems integrated with a chemistry analyzer.
Reason for recall
It has been determined that, in some cases, pulley drive components have broken parts or have become loose, mispositioned, or disconnected from their drive shafts. When these defects occur, the pulley may not turn at all, or may turn at an incorrect speed, resulting in inadequate washing or mixing of the contents of the reaction vessel.