Recall Z-0599-2019— Tosoh Bioscience Inc
Class II · Terminated
Class II FDA medical device recall by Tosoh Bioscience Inc, initiated 2018-09-28, terminated 2020-09-03. FDA's recall record names no specific 510(k) clearance for this event. Product: AIA-2000 Analyzer, Product code 022100, 022100R, 022101, 022101R The AIA-2000 analyzer and AIA-900 analyzer are capable of performing three methods of immunoassay: an immunoenzymetric (IEMA) or sandwich immunoassay, a competitive binding (EIA) immunoassay, and a two-step immunoenzymetric immunoassay.. Reason: Tosoh Bioscience has become aware of a potential issue with the AIA-900 and AIA-2000 analyzers that could result in slippage of the incubator unit contained within the system. If the tension of the timing belt is not adjusted correctly, it is possible for a misalignment that will result in a mechanical error. Tosoh has taken the required actions to inspect the tension of the timing belt during preventative maintenance and correct the instruments where this issue has been observed. If incubator slippage occurs, the unit will cease to function, and a system error will be reported. If this occurs, there may be a delay in testing and reporting patient test results. There have been no reports of injury related to this issue..
- Recalling firm
- Tosoh Bioscience Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-09-28
- Terminated
- 2020-09-03
- Firm location
- Grove City, OH, United States
Product description
AIA-2000 Analyzer, Product code 022100, 022100R, 022101, 022101R The AIA-2000 analyzer and AIA-900 analyzer are capable of performing three methods of immunoassay: an immunoenzymetric (IEMA) or sandwich immunoassay, a competitive binding (EIA) immunoassay, and a two-step immunoenzymetric immunoassay.
Reason for recall
Tosoh Bioscience has become aware of a potential issue with the AIA-900 and AIA-2000 analyzers that could result in slippage of the incubator unit contained within the system. If the tension of the timing belt is not adjusted correctly, it is possible for a misalignment that will result in a mechanical error. Tosoh has taken the required actions to inspect the tension of the timing belt during preventative maintenance and correct the instruments where this issue has been observed. If incubator slippage occurs, the unit will cease to function, and a system error will be reported. If this occurs, there may be a delay in testing and reporting patient test results. There have been no reports of injury related to this issue.
Data sourced from openFDA. This site is unofficial and independent of the FDA.