Recall Z-0597-2019— COVIDIEN LLC
Class II · Terminated
Class II FDA medical device recall by COVIDIEN LLC, initiated 2018-10-31, terminated 2020-05-26. It names one FDA 510(k) clearance: K810156. Product: Salem Sump Dual Lumen Stomach Tube 10 Fr/Ch (3.3 mm) x 36" (91 cm), Product Code 8888264911 Product Usage: The Salem Sump Tubes are intended for decompression and drainage of stomach contents following injury, illness, or surgery in patients with intestinal obstruction, and to prevent the distention of the stomach.. Reason: This voluntary recall is being conducted due to incorrect packaging. In one lot of Salem Sump Dual Lumen Stomach Tubes the 10Fr product was packaged in a 16Fr pouch. The use of a 10 Fr Salem Sump Tube when a 16Fr is needed could result in decreased fluid collection, with the potential to affect patient care, particularly in an emergency situation where the correct size is not readily available..
- Recalling firm
- COVIDIEN LLC
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-10-31
- Terminated
- 2020-05-26
- Firm location
- Mansfield, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Salem Sump Dual Lumen Stomach Tube 10 Fr/Ch (3.3 mm) x 36" (91 cm), Product Code 8888264911 Product Usage: The Salem Sump Tubes are intended for decompression and drainage of stomach contents following injury, illness, or surgery in patients with intestinal obstruction, and to prevent the distention of the stomach.
Reason for recall
This voluntary recall is being conducted due to incorrect packaging. In one lot of Salem Sump Dual Lumen Stomach Tubes the 10Fr product was packaged in a 16Fr pouch. The use of a 10 Fr Salem Sump Tube when a 16Fr is needed could result in decreased fluid collection, with the potential to affect patient care, particularly in an emergency situation where the correct size is not readily available.
Data sourced from openFDA. This site is unofficial and independent of the FDA.