Recall Z-0591-2024— Insulet Corporation
Class I · Ongoing
Class I FDA medical device recall by Insulet Corporation, initiated 2023-11-30. It names one FDA 510(k) clearance: K231826. Product: Omnipod 5 App (on compatible Android smartphones), Software Versions 1.1 - 1.2.3 (current). US: PT-000599- Omnipod 5 App Software Package (provided through Google Play), EU: PT-001323- Omnipod 5 App Software Package (provided on locked-down Controllers (PT-001194 and PT-001193). Used to control an insulin pump.. Reason: The bolus calculator is not recording the decimal point if it is the first value entered when changing a bolus dose. This may lead to an over-delivery of insulin to the user if the user does not recognize the error on the bolus calculator screen or the confirmation screen prior to starting the bolus..
- Recalling firm
- Insulet Corporation
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2023-11-30
- Firm location
- Acton, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Omnipod 5 App (on compatible Android smartphones), Software Versions 1.1 - 1.2.3 (current). US: PT-000599- Omnipod 5 App Software Package (provided through Google Play), EU: PT-001323- Omnipod 5 App Software Package (provided on locked-down Controllers (PT-001194 and PT-001193). Used to control an insulin pump.
Reason for recall
The bolus calculator is not recording the decimal point if it is the first value entered when changing a bolus dose. This may lead to an over-delivery of insulin to the user if the user does not recognize the error on the bolus calculator screen or the confirmation screen prior to starting the bolus.
Data sourced from openFDA. This site is unofficial and independent of the FDA.