Recall Z-0589-2022— Microbiologics Inc

Class II · Terminated

Class II FDA medical device recall by Microbiologics Inc, initiated 2020-02-18, terminated 2022-07-11. It names one FDA 510(k) clearance: K861022. Product: QC Sets and Panels, KWIK-STIK Blood Culture Identification (BCID) Verification Panel (Live Culture). KWIK-STIK kit that contains Catalog Number 0959P. In Vitro Diagnostic Control microorganisms. Reason: Tests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release..

Recalling firm
Microbiologics Inc
Classification
Class II
Status
Terminated
Initiated
2020-02-18
Terminated
2022-07-11
Firm location
Saint Cloud, MN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

QC Sets and Panels, KWIK-STIK Blood Culture Identification (BCID) Verification Panel (Live Culture). KWIK-STIK kit that contains Catalog Number 0959P. In Vitro Diagnostic Control microorganisms

Reason for recall

Tests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release.

Data sourced from openFDA. This site is unofficial and independent of the FDA.