Recall Z-0588-2026— Pivotal Health Solutions, Inc.

Class II · Ongoing

Class II FDA medical device recall by Pivotal Health Solutions, Inc., initiated 2025-10-09. FDA's recall record names no specific 510(k) clearance for this event. Product: Quantum Intersegmental Table, Model/Catalog Number: AMQ400H; heated massage table. Reason: The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl cushion..

Recalling firm
Pivotal Health Solutions, Inc.
Classification
Class II
Status
Ongoing
Initiated
2025-10-09
Firm location
Watertown, SD, United States

Product description

Quantum Intersegmental Table, Model/Catalog Number: AMQ400H; heated massage table

Reason for recall

The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl cushion.