Recall Z-0587-2017— Inpeco S.A.
Class II · Terminated
Class II FDA medical device recall by Inpeco S.A., initiated 2016-10-04, terminated 2019-05-31. It names one FDA 510(k) clearance: K121012. Product: FlexLab, Accelerator a3600 and Aptio Automation are modular systems designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.. Reason: Inpeco received several notifications from the field about burnt schuko sockets. The customers reported that some smoke was seen coming from the damaged sockets..
- Recalling firm
- Inpeco S.A.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-10-04
- Terminated
- 2019-05-31
- Firm location
- Lugano, N/A, Switzerland
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
FlexLab, Accelerator a3600 and Aptio Automation are modular systems designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.
Reason for recall
Inpeco received several notifications from the field about burnt schuko sockets. The customers reported that some smoke was seen coming from the damaged sockets.