Recall Z-0586-2022— Microbiologics Inc
Class II · Terminated
Class II FDA medical device recall by Microbiologics Inc, initiated 2020-02-18, terminated 2022-07-11. It names one FDA 510(k) clearance: K861022. Product: LYFO-DISK Enterococcus faecalis derived from ATCC 51299 In Vitro Diagnostic Control microorganisms. Reason: Tests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release..
- Recalling firm
- Microbiologics Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2020-02-18
- Terminated
- 2022-07-11
- Firm location
- Saint Cloud, MN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
LYFO-DISK Enterococcus faecalis derived from ATCC 51299 In Vitro Diagnostic Control microorganisms
Reason for recall
Tests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release.