Recall Z-0585-2022— Microbiologics Inc

Class II · Terminated

Class II FDA medical device recall by Microbiologics Inc, initiated 2020-02-18, terminated 2022-07-11. It names one FDA 510(k) clearance: K861022. Product: KWIK-STIK, Quality control kit for culture media, sold as: a. KWIK-STIK 2-pack Enterococcus faecalis derived from ATCC 51299 b. KWIK-STIK 6-pack Enterococcus faecalis derived from ATCC 51299. Reason: Tests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release..

Recalling firm
Microbiologics Inc
Classification
Class II
Status
Terminated
Initiated
2020-02-18
Terminated
2022-07-11
Firm location
Saint Cloud, MN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

KWIK-STIK, Quality control kit for culture media, sold as: a. KWIK-STIK 2-pack Enterococcus faecalis derived from ATCC 51299 b. KWIK-STIK 6-pack Enterococcus faecalis derived from ATCC 51299

Reason for recall

Tests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release.

Data sourced from openFDA. This site is unofficial and independent of the FDA.