Recall Z-0585-2015— Heart Sync, Inc.

Class I · Terminated

Class I FDA medical device recall by Heart Sync, Inc., initiated 2014-11-11, terminated 2015-11-24. It names one FDA 510(k) clearance: K080421. Product: ADULT Radiotranslucent Electrode, Part number T100LO-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart Electrodes Plug Style Connector", Intended for use during defibrillation, cardioversion, pacing, and ECG monitoring, Single use.. Reason: Heart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads are not connected to the unit prior to use. The FR3, however, does not require pre-connection and the user may not discover the incompatibility issue until the.

Recalling firm
Heart Sync, Inc.
Classification
Class I
Status
Terminated
Initiated
2014-11-11
Terminated
2015-11-24
Firm location
Ann Arbor, MI, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ADULT Radiotranslucent Electrode, Part number T100LO-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart Electrodes Plug Style Connector", Intended for use during defibrillation, cardioversion, pacing, and ECG monitoring, Single use.

Reason for recall

Heart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads are not connected to the unit prior to use. The FR3, however, does not require pre-connection and the user may not discover the incompatibility issue until the

Data sourced from openFDA. This site is unofficial and independent of the FDA.