Recall Z-0583-2020— GN Hearing A/S
Class II · Terminated
Class II FDA medical device recall by GN Hearing A/S, initiated 2019-09-18, terminated 2023-05-31. FDA's recall record names no specific 510(k) clearance for this event. Product: Behind-the-Ear hearing aids, Model Numbers: Q962-DRW, T561-DRW, T562-DRW, T762-DRW, T777-DW, AMZ964-DRW, ET588-DWH, ET598-DW, ET788-DWH, ET798-DW, ET988-DWH, ET998-DW, FT861-DRW, FT862-DRW, RE561-DRWC, RE762-DRW, RE961-DRWC, RE962-DRW, PZ862-DRW, LT561-DRW, LT561-DRWZ, LT562-DRW, LT588-DW, LT761-DRW, LT761-DRWZ, LT762-DRW, LT777-DW, LT788-DW, LT961-DRW, LT961-DRWZ, LT962-DRW, LT967-DW, LT977-DW, LT988-DW, TST1763-DRW, TST1764-DRW, TST663-DRW, TST963-DRW, TST964-DRW, VA861-DRW, VA861-DRWZ, VA862-DRW. Reason: GN Hearing received reports from our manufacturing site in US regarding a software bug in our service & repair tool which affected the capture and restoration of user data in a limited number of repaired hearing aids. The software bug caused these repaired hearing aids to contain altered specifications for sound amplification, potentially putting users residual hearing at risk of experiencing unintended additional available gain to varying degrees i.a.18dB flat linear gain or potentially in worst case Full On Gain. However, no injuries have been reported..
- Recalling firm
- GN Hearing A/S
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-09-18
- Terminated
- 2023-05-31
- Firm location
- Ballerup, Denmark
Product description
Behind-the-Ear hearing aids, Model Numbers: Q962-DRW, T561-DRW, T562-DRW, T762-DRW, T777-DW, AMZ964-DRW, ET588-DWH, ET598-DW, ET788-DWH, ET798-DW, ET988-DWH, ET998-DW, FT861-DRW, FT862-DRW, RE561-DRWC, RE762-DRW, RE961-DRWC, RE962-DRW, PZ862-DRW, LT561-DRW, LT561-DRWZ, LT562-DRW, LT588-DW, LT761-DRW, LT761-DRWZ, LT762-DRW, LT777-DW, LT788-DW, LT961-DRW, LT961-DRWZ, LT962-DRW, LT967-DW, LT977-DW, LT988-DW, TST1763-DRW, TST1764-DRW, TST663-DRW, TST963-DRW, TST964-DRW, VA861-DRW, VA861-DRWZ, VA862-DRW
Reason for recall
GN Hearing received reports from our manufacturing site in US regarding a software bug in our service & repair tool which affected the capture and restoration of user data in a limited number of repaired hearing aids. The software bug caused these repaired hearing aids to contain altered specifications for sound amplification, potentially putting users residual hearing at risk of experiencing unintended additional available gain to varying degrees i.a.18dB flat linear gain or potentially in worst case Full On Gain. However, no injuries have been reported.
Data sourced from openFDA. This site is unofficial and independent of the FDA.