Recall Z-0575-2024— Karl Storz Endoscopy

Class II · Ongoing

Class II FDA medical device recall by Karl Storz Endoscopy, initiated 2023-10-24. FDA's recall record names no specific 510(k) clearance for this event. Product: Injection Needle, curved 23 cm, REF: 8598A; Injection Needle, straight, 23 cm, REF: 8598B; Injection Needle, distal 45-degrees, 23 cm, REF: 8598D. Reason: The efficacy of the manual cleaning process cannot be assured for the affected injection needles intended for general surgery and visceral surgery, which if used, may expose the patient to a higher risk of infection..

Recalling firm
Karl Storz Endoscopy
Classification
Class II
Status
Ongoing
Initiated
2023-10-24
Firm location
El Segundo, CA, United States

Product description

Injection Needle, curved 23 cm, REF: 8598A; Injection Needle, straight, 23 cm, REF: 8598B; Injection Needle, distal 45-degrees, 23 cm, REF: 8598D

Reason for recall

The efficacy of the manual cleaning process cannot be assured for the affected injection needles intended for general surgery and visceral surgery, which if used, may expose the patient to a higher risk of infection.