Recall Z-0566-2016— Aesculap, Inc.
Class II · Terminated
Class II FDA medical device recall by Aesculap, Inc., initiated 2015-11-17, terminated 2016-10-13. FDA's recall record names no specific 510(k) clearance for this event. Product: Valve XS Atrium Retractor FC429R 52154756 The Valve XS Atrium Lift Retractor is used in heart surgery. The retractor is used for the temporary dilation of the left atrium in mitral valve surgery and for assistance in tricuspid valve surgery.. Reason: AAG received a customer complaint that corrosion had been found by a hospitals central sterile service department on the ball joint of an FC424R blade. Investigation revealed that the corrosion was due to incorrect material used in the manufacturing of the disk spring of the ball joint by the supplier.
- Recalling firm
- Aesculap, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2015-11-17
- Terminated
- 2016-10-13
- Firm location
- Center Valley, PA, United States
Product description
Valve XS Atrium Retractor FC429R 52154756 The Valve XS Atrium Lift Retractor is used in heart surgery. The retractor is used for the temporary dilation of the left atrium in mitral valve surgery and for assistance in tricuspid valve surgery.
Reason for recall
AAG received a customer complaint that corrosion had been found by a hospitals central sterile service department on the ball joint of an FC424R blade. Investigation revealed that the corrosion was due to incorrect material used in the manufacturing of the disk spring of the ball joint by the supplier