Recall Z-0565-2013— Del Mar Reynolds Medical, Ltd.
Class II · Terminated
Class II FDA medical device recall by Del Mar Reynolds Medical, Ltd., initiated 2012-11-02, terminated 2013-07-11. It names one FDA 510(k) clearance: K003251. Product: BleaseFocus Anesthesia Workstation and 700/900 Series Ventilators are anesthesia gas machine. In most cases, the customer ordered both a BleaseFocus anesthesia workstation and a 700/900 ventilator. In some cases, the customer may order a 700/900 stand-alone ventilator would could be used with a BleaseFocus anesthesia workstation at a later date.. Reason: The BleaseFocus Anesthesia Workstation and 700/900 Series Ventilators were shipped with manuals that describe a secondary O2 Supply Failure alarm which has not been incorporated into the system. The manuals were prematurely updated for a feature that is not yet available..
- Recalling firm
- Del Mar Reynolds Medical, Ltd.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2012-11-02
- Terminated
- 2013-07-11
- Firm location
- Hertford, United Kingdom
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
BleaseFocus Anesthesia Workstation and 700/900 Series Ventilators are anesthesia gas machine. In most cases, the customer ordered both a BleaseFocus anesthesia workstation and a 700/900 ventilator. In some cases, the customer may order a 700/900 stand-alone ventilator would could be used with a BleaseFocus anesthesia workstation at a later date.
Reason for recall
The BleaseFocus Anesthesia Workstation and 700/900 Series Ventilators were shipped with manuals that describe a secondary O2 Supply Failure alarm which has not been incorporated into the system. The manuals were prematurely updated for a feature that is not yet available.